80kVp低管电压扫描联合全模型迭代重建冠状动脉CTA成像的临床应用 被引量:8

Clinical application of coronary CTA at 80kVp combined with iterative model reconstruction
下载PDF
导出
摘要 目的探讨80kVp低管电压扫描联合全模型迭代重建(IMR)冠状动脉CTA成像的可行性。方法 256层螺旋CT冠状动脉CTA扫描BMI为20~25 kg/m^2的患者60例;其中30例患者行80kVp扫描采用IMR重建(A组),30例患者行100kVp扫描采用iDose^4算法重建(B组)。采用定量及定性方法分别对两组图像质量进行评价,并对两组的图像质量及辐射剂量进行对比分析,定量图像质量评价指标包括冠脉主要分支近、中、远段CT值、图像噪声及对比噪声比(CNR);定性图像质量评价指标包括血管边缘锐利度、条纹伪影及总体图像质量,采用5分法。结果 80kVp扫描较100kVp有效辐射剂量降低45.3%。A组冠脉各节段CT值及CNR均明显高于B组(P<0.05)。A组冠脉各节段图像噪声均低于B组。A组各定性图像质量指标评分均明显高于B组(P<0.05)。结论对于BMI为20~25 kg/m^2的患者,80kVp低管电压冠状动脉CTA扫描联合IMR可实现低剂量冠状动脉成像。 Objective To investigate the feasibility of coronary computed tomographic angiography(CTA) at 80 kVp combined with iterative model reconstruction(IMR). Methods Sixty patients with BMI of 20-25 kg/m^2 underwent 256-slice coronary CTA. Among them, 30 patients were subjected to 80 kVp voltage scanning, followed by algorithm of IMR(group A) and 30 patients were screened with 100 kVp voltage scanning combined with iDose4(group B). Quantitative and qualitative indicators were used to evaluate the image quality of the two groups. The quantitative indicators included CT value, images noise, and compared noise ratio(CNR) of the proximal, middle and distal part of the coronary’s main branch. The qualitative indexes included vascular edge sharpness, streak artifacts and image quality, all of which were evaluated by a 5-point system. Results The effective radiation dose of 80 kVp decreased by 45.3% compared with 100 kVp. The CT value and CNR in each segment of coronary artery in group A were significantly higher than those in group B(P<0.05). The image noise of each segment of coronary artery in group A was lower than that in group B. The qualitative image quality index scores of group A were significantly higher than those of group B(P<0.05). Conclusion For patients with a BMI of 20-25 kg/m^2, Low-dose coronary artery imaging can be achieved by 80 kVp low-tube voltage coronary CTA combined with IMR.
作者 赵正凯 周瀚 程绍玲 赖声远 曲晓峰 陈宏海 ZHAO Zhengkai;ZHOU Han;CHENG Shaoling;LAI Shengyuan;QU Xiaofeng;CHEN Honghai(Department of Radiology,the Third People’s Hospital of Chengdu,Chengdu 610031,China;Department of Radiology,the Second Affiliated Hospital of Dalian Medical University,Dalian 116027,China)
出处 《大连医科大学学报》 CAS 2019年第6期506-510,共5页 Journal of Dalian Medical University
关键词 全模型迭代算法 低管电压 冠状动脉CTA成像 iterative model reconstruction(IMR) low voltage scanning coronary computed tomography angiography(CCTA)
  • 相关文献

参考文献1

二级参考文献13

  • 1Brenner D J, Hall EJ. Computed tomography- an increasing source of radiation exposure[J].N Engl J Med, 2007, 357(22): 2277-2284. 被引量:1
  • 2Prakash P, Kalra MK, Digumarthy SR, et al. Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience[J].J Comput Assist Tomogx, 2010, 34(1):40-45. 被引量:1
  • 3Schindera ST, Diedrichsen L,MtillerHC, et al. Iterative reconstruction algorithm for abdominal muhidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study[J]. Radiology, 2011,260(2):454-462. 被引量:1
  • 4Funama Y, Taguchi K, Utsunomiya D, et al. Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography[J].J Comput Assist Tomogr, 2011, 35(4):480-485. 被引量:1
  • 5Cheng Z, Wang X, Duan Y, et al. Low-dose prospective ECG- triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience[J].Eur Radiol, 2010, 20(10):2503-2511. 被引量:1
  • 6Paul JF, Rohnean A, Elfassy E, et al. Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual - source machine in infants and small children with congenital heart disease[J].Pediatr Radiol, 2011,41 (2):244-249. 被引量:1
  • 7Tsai IC, Chen MC, Jan SL, et al. Neonatal cardiac multidetector row CT: why and how we do it[J].Pediatr Radiol, 2008, 38(4):438-451. 被引量:1
  • 8Huang MP, Liang CH, Zhao Z J, et al. Evaluation of image quality and radiation dose at prospective ECG-triggered axial 256-slice multi-detector CT in infants with congenital heart disease[J].Pediatr Radiol, 2011, 41(7):858-866. 被引量:1
  • 9Srichai MB, Lira RP, Donnino R, et al. Low-dose, prospective triggered high- pitch spiral coronary computed tomography angiography: comparison with retrospective spiral technique [J].Acad Radiol, 2012, 19(5):554-561. 被引量:1
  • 10Chang W, Lee JM, Lee K, et al. Assessment of a model- based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomozraDhvHl.Invest Radiol. 2013.48(8):598-606. 被引量:1

共引文献32

同被引文献68

引证文献8

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部