The Effect of CARE- kv on Radiation Dose and Image Quality of Dual Source Coronary CT Angiography
原文传递
导出
摘要 目的探讨双源CT的CARE-kV扫描技术对冠状动脉成像(CTCA)的辐射剂量及图像质量的影响。方法回顾性分析行CTCA检查的患者393例,根据管电压设置方法分为两组:A组189例,应用CARE-kV技术智能选择管电压;B组204例,根据患者的体重指数(BMI)确定管电压,两组均应用CARE Dose 4D技术自动选择管电流。测量并记录主动脉根部及左、右冠状动脉主干的CT值及其标准差(SD),并进一步计算对比噪声比(CNR)和信噪比(SNR)值,以用于客观图像质量评估。然后由两名医师通过5分评价法对两组图像质量进行主观评价,取其均值纳入统计学分析。比较两组CNR,SNR,CT剂量指数(CTDIvol),剂量长度乘积(DLP)。结果 A组客观图像质量优于B组(P<0.05),主观图像质量差异无统计学差异(P=2.18);A组较B组CTDIvol降低32.4%[(20.08±18.20)mGy vs(26.49±21.55)mGy,P=0.003],DLP降低22.93%[(279.88±268.55)mGy·cm vs(363.15±297.0)mGy·cm,P=0.043],且A组较B组明显增加了80 kV管电压的应用率(χ2=73.289,P<0.05)。结论应用CARE-kV智能最佳kV扫描技术进行冠状动脉成像,不仅能降低受检者接受的辐射剂量,而且能同时提高图像质量。 Objective To discuss effect of CARE kv on radiation dose and image quality of dual source coronary CT angiography(CTA). Methods Consecutive patients (393 patients) were respectively enrolled: Group A (189 patients) underwent dual source coronary CTA with CARE kv and CARE Dose 4D, in whom tube potential and tube current were selected automatically ; group B (204 patients) underwent CTA with a BMI based tube potential and CARE Dose 4D. Quantitative image quality parameters, including image noise, attenuation, CNR and SNR, the CT density in a region of in terest placed in the aortic root, LM and RCA were measured. The scores of imaging quality were evaluated by two experi enced reviewers and CNR, SNR,CTDIvol, DLP were compared between the two group. Results The value of CNR and SNR in group A were significantly higher than that in group B(P 〈0.05). The value of CTDIvol in group A were 32.4% significantly lower than that in group B( t = 0. 098, P = 0. 003 ), DLP were 22.93% significantly lower( t = 4. 838, P = 0.043) with a significantly more frequent use of 80 kV ( x2 = 73. 289,P 〈 0.05). Conclusion The application of CARE kv in dual source coronary CT angiography can significantly reduce radiation dose and improve image quality.
出处 《临床放射学杂志》 CSCD 北大核心 2013年第11期1645-1649,共5页 Journal of Clinical Radiology
关键词 冠状动脉体层摄影术管电压辐射剂量 图像质量 Coronary angiography Tomography Tube potential Radiation dose Image quality
  • 相关文献

参考文献2

二级参考文献10

  • 1Slovis TL.CT and computed radiography:the pictures aregreat,but is the radiation dose greater than required?[J].Am JRoentgenol,2002,179(1):39-41. 被引量:1
  • 2Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiationdose optimization[J].Radiology.2004,230(3):619-628. 被引量:1
  • 3Brix G,Lechel U,Petersheim M,et al.Dynamic contrast-enhancedCT studies:balancing patient exposure and image noise[J].InvestRadiol.2011,46(1):64-70. 被引量:1
  • 4Yu L,Li H,Fletcher JG,et al.Automatic selection of tube potentialfor radiation dose reduction in CT:a general strategy[J].MedPhys,2010,37(1):234-243. 被引量:1
  • 5Kalender WA,Deak P,Kellermeier M,et al.Application-andpatient size-dependent optimization of X-ray spectra for CT[J].Med Phys,2009,36(3):993-1007. 被引量:1
  • 6Alkadhi H,Schindera ST.State of the art low-dose CTangiography of the body[J].Eur J Radiol,2011,80(1):36-40. 被引量:1
  • 7Winklehner A,Goetti R,Baumueller S,et al.Automatedattenuation-based tube potential selection for thoracoabdominalcomputed tomography angiography:improved doseeffectiveness[J].Invest Radiol,2011,46(12):767-773. 被引量:1
  • 8CAREkV(computer program).Version syngo CT 2011 A(VA40)[M].Forchheim,Germany:Siemens Healthcare,2011. 被引量:1
  • 9Alibek S,Brand M,Suess C,et al.Dose reduction in pediatriccomputed tomography with automated exposure control[J].AcadRadiol2011,Jun,18(6):690-693. 被引量:1
  • 10Smith AB,Dillon WP,Gould R,et al.Radiation dose-reductionstrategies for neuroradiology CT protocols[J].AJNR Am J Neuroradiol,2007,28(9):1628-1632. 被引量:1

共引文献94

同被引文献42

  • 1高建华,孙宪昶,李剑颖,王贵生,李娜,戴汝平.后置滤过器C_2对64层螺旋CT冠状动脉成像质量的影响[J].中国医学影像技术,2007,23(1):73-76. 被引量:31
  • 2Winklehner A,Goetti R, Baumueller S, et al. Automated attenua- tion based tube potential selection for thoraeoahdominal computed tomography angiography.- improved dose effectiveness[J]. Invest Radiol,2011,46(12) :767 773. 被引量:1
  • 3Aehenbaeh S,Goroll T,Seltmann M,et al. Detection of coronary ar- tery stenoses by low-dose, prospectinely ECG-triggered, high-pitch spiral coronary CT angiography [ J ]. JACC Careiovasc Imaging, 2011,4:328-337. 被引量:1
  • 4Becker CR, Knez A, Ohnesorge B, et al. Imaging of noncalcified cor- onary plaques using helical CT with restrospective ECG gating. A JR, 2000,175:423-424. 被引量:1
  • 5Brenner DJ , Hall [J]. Computed tomography-an increasing source of radiation exposure[ J]. N Engl J Med ,2007,357:2277-2284. 被引量:1
  • 6Einstein A J, Henzlova M J, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomo- graphy coronary angiography [ J ]. JAMA,2007,298 : 317 -323. 被引量:1
  • 7Hundt W, Rust FP, Stabler AM, et al. Dose reduction in muhislice computed tomogTaphy[ J ]. J Comput Assist Tomogr, 2005,29 : 140- 147. 被引量:1
  • 8Liistro F, Falsinni G, Bolognese L. The clinical burden of contrast media-induced nephropathy [ J ]. Ital Heart J, 2003,4 : 668 -676. 被引量:1
  • 9Cademartiri F, Maffei E, Palumbo AA, et al. Influence of intra-coro- nary enhancement on diagnostic accuracy with 64-slice CT coronary angiography. Eur Radio1,2008 ,18 :576-83. 被引量:1
  • 10Winklehner A, Karlo C, Puippe G, et al. Raw data-based iterative reconstruction in body CTA:evaluation of radiation dose saving po- tential [ J ]. Eur Radio1,2011,21:2521-2526. 被引量:1

引证文献7

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部