期刊文献+

原子荧光法测定地表水中砷的影响因素讨论 被引量:10

Discussion of Influence Factors in Measuring Arsenic of Surface Water with Atom Fluorescence
下载PDF
导出
摘要 应用氢化物-原子荧光光谱法测定地表水中的砷含量,对测定过程中酸介质的纯度、还原剂用量、样品的消解等影响因素进行讨论。通过优化酸纯度和还原剂的用量,选择优级纯酸及质量浓度为2%的硼氢化钾,在此条件下,所做标准曲线的相关系数为0.999 8,检出限为0.015 9 ug/L,相对标准偏差为0.73%。 This paper discusses the arsenic content measurement of surface water with Hydride Generation - Atomic Fluorescence Spectrometry ( HG - AFS) and the influence factors concerning purity of acidic medium, the amount of reducing agent and sample digestion during the measurement process. By optimi- zing purity of acidic medium, proper use of reducing agent, choosing high quality acid and 2% potassium borohydride, the correlation coefficient of the standard curve comes to 0. 999 8 and the detection limit is 0. 015 9 ug/L, with a relative standard deviation of 0. 73%.
出处 《昆明冶金高等专科学校学报》 CAS 2012年第1期73-76,共4页 Journal of Kunming Metallurgy College
关键词 原子荧光 光谱法 影响因素 atom fluorescence spectroscopic methods arsenic influence factors
  • 相关文献

参考文献6

  • 1梁红.环境监测[M]武汉:武汉理工大学出版社,2008. 被引量:1
  • 2刘明钟;汤志勇.原子荧光光谱分析[M]北京:化学工业出版社,2007. 被引量:1
  • 3国家环境保护总局.水和废水监测分析方法[M]北京:中国环境科学出版社,2002. 被引量:1
  • 4高红武;周清.应用化学[M]北京:中国环境科学出版社,2011. 被引量:1
  • 5黄一石;吴朝华;杨小林.仪器分析[M]北京:化学工业出版社,2008. 被引量:1
  • 6赵联朝;董正荣.分析仪器及维修[M]北京:中国环境科学出版社,2007. 被引量:1