吸入用乙酰半胱氨酸溶液雾化特性的体外评价 被引量:7

In vitro evaluation of nebulization characteristics of acetylcysteine inhalation solution
原文传递
导出
摘要 目的研究吸入用乙酰半胱氨酸溶液的雾化特性,建立体外评价方法。方法使用呼吸模拟装置对吸入用乙酰半胱氨酸溶液的递送速率和递送总量进行研究,使用新一代药用多级撞击器(NGI)测定空气动力学粒径分布。结果不同雾化装置检测递送速率和递送总量数据存在明显差异,但使用同品牌雾化装置时吸入用乙酰半胱氨酸溶液自制样品和参比制剂的递送速率和递送总量无明显差异。不同雾化装置检测空气动力学粒径分布数据存在明显差异,但是使用同一品牌雾化装置,吸入用乙酰半胱氨酸溶液自制样品与参比制剂无明显差异。结论所测定的体外数据描述了吸入用乙酰半胱氨酸溶液的体外特征,可为该制剂的体外评价提供参考。 Objective To study the nebulization characteristics and establish in vitro evaluation methods of acetylcysteine inhalation solution.Methods The delivery rate and total delivery amount of acetylcysteine inhalation solution were measured by breathing simulator,and aerodynamic particle size distribution was determined by next generation pharmaceutical impactor(NGI).Results There were significant differences in the delivery rate and total delivery amount among different atomizing devices,but there was no significant difference in the delivery rate and total delivery amount between acetylcysteine inhalation solution self-made and the reference preparation when using the same brand atomization device.There were significant differences in the aerodynamic particle size distribution data measured by different atomizing devices,but there was no significant difference between the acetylcysteine inhalation solution self-made sample and the reference preparation with the same brand atomization device.Conclusion The in vitro characterization of acetylcysteine inhalation solution can be described by these methods,which can provide reference for the in vitro evaluation of acetylcysteine inhalation solution.
作者 王婷 魏焕莉 周晓雪 胡杰 WANG Ting;WEI Huan-li;ZHOU Xiao-xue;HU Jie(Tianjin Key Laboratory of Pulmonary Drug Delivery Key Technology Enterprise,Tianjin 300385,China)
出处 《现代药物与临床》 CAS 2020年第12期2317-2320,共4页 Drugs & Clinic
关键词 吸入用乙酰半胱氨酸溶液 雾化特性 递送速率 递送总量 空气动力学粒径分布 acetylcysteine inhalation solution nebulization characteristics delivery rate delivered amount aerodynamic particle size distribution
  • 相关文献

参考文献3

二级参考文献13

  • 1Weda M,Zanen P,de Boer AH,et al.Equivalence testing of salbutamol dry powder inhalers:in vitro impaction results versus in vivo efficacy[J].Int J Pharm,2002,249:247-255. 被引量:1
  • 2Ashurst IC,Herman CS,Li-Bovet L,et al.Metered dose inhaler for albuterol:US,6131566[P].2000-10-17. 被引量:1
  • 3Dunbar C,Mitchell J.Analysis of cascade impactor mass distributions[J].J Aerosol Med,2005,18(4):439-451. 被引量:1
  • 4Stein SW.Size distribution measurements of metered dose inhalers using Andersen Mark Ⅱ cascade impactors[J].Int J Pharm,1999,186:43-52. 被引量:1
  • 5de Boer AH,Gjaltema D,Hagedoorn P,et al.Characterization of inhalation aerosols:A critical evaluation of cascade impactor and laser diffraction technique[J].Int J Pharm,2002,249(1-2):219-231. 被引量:1
  • 6Harris JA,Stein SW,Myrdal PB.Evaluation of the TSI aerosol impactor 3306/3321 system using a redesigned impactor stage with solution and suspension metered-dose inhalers[J].AAPS Pharm Sci Tech,2006,7(1):Article 20.(http://www.aapspharmscitech.org) 被引量:1
  • 7Mitchell JP,Nagel MW Time-of-flight aerodynamic particle size analyzers:Their use and limitations for the evaluation of medical aerosols[J].J Aerosol Med,1999,12(4):217-240. 被引量:1
  • 8Oskouie AK,Noll KE,Wang H-C.Minimizing the effect of density in determination of particle aerodynamic diameter using a time of flight instrument[J].Jaerosol Sci,2003,34:501-506. 被引量:1
  • 9Clark A,Borgstrom L.In vitro testing of pharmaceutical aerosols and predicting lung deposition from in vitro measurements[M].In:H.Bisgaard,C.O'Callaghan,and G.C.Smaldone.Eds,Drug Delivery to the Lung.Marcel Dekker,NY,USA,2002:105-142. 被引量:1
  • 10LeBelle M,Pike RK,Graham SJ,et al.Metered-dose inhalers I:drug content and particle size distribution of beclomethasone dipropionate[J].J Pharm Biomed Anal,1996,14:793-800. 被引量:1

共引文献33

同被引文献64

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部