期刊文献+

迈瑞D3除颤监护仪常见故障分析与维修 被引量:5

Failure Analysis and Maintenance of Mindray D3 Defibrillator Monitor
下载PDF
导出
摘要 根据我院故障数据统计,D3除颤仪临床故障主要集中在前壳模块、多参数模块和电源模块,故障临床主要表现为无法开机、无法充放电、黑屏无显示、无心电波形,此四类故障发生率约占故障总量的64%。故障的快速排查与解决不仅能保障设备的完好率,也可以降低临床使用成本,同时确保临床使用安全。 According to the statistics of the fault data of our hospital,the clinical failures of D3 defibrillator were mainly concentrated in the front shell module,multi-parameter module and power module.The main clinical manifestations of failures were failure to boot,failure to charge and discharge,no display on black screen,and no ECG waveform,which these four types accounted for about 64%of total failures.Rapid troubleshooting and maintenance can not only ensure the intact rate of defibrillator,but also reduce maintenance costs,while effectively ensuring the safety of equipment.
作者 魏昊业 宗会迁 柳青 杨存 WEI Haoye;ZONG Huiqian;LIU Qing;YANG Cun(Department of Equipment Maintenance,The Second Hospital of Hebei Medical University,Shijiazhuang Hebei 050000,China)
出处 《中国医疗设备》 2020年第9期173-175,共3页 China Medical Devices
关键词 迈瑞D3除颤仪 电源模块 高压治疗模块 多参数模块 Mindray D3 defibrillation monitor power module high voltage treatment module multi-parameter module