摘要
目的:运用除颤分析仪对心脏除颤器释放能量性能进行检测,揭示其中变化规律并进行科学论证,使检测数据在设备的维护保障工作中发挥作用。方法:使用FLUKE IMPULSE 7000DP除颤器分析仪,遵循国家计量技术规范JJF1149-2014心脏除颤器校准规范,运用图、表综合的分析方法对检测数据进行分析。结果:通过对释放能量检测数据的分析,发现释放能量变化规律及检测数据分布差异现象,对现象原因或影响因素进行评估,通过性能检测筛查出释放能量超差设备。结论:通过心脏除颤器释放能量检测及数据分析,掌握了在用设备整体的性能状态,提高对除颤器分析仪的技术认知及应用水平,避免了临床诊治过程中由于设备性能隐患带来的医疗风险。
Objective To adopt defibrillation analyzer to detect performance of delivered energy of cardiac defibrillator and to reveal its relevant changing regular, and carry out scientific verification so as to take the detection data to play a role in the maintenance work of equipment. Methods: According to National Metrological technical specification “JJF 1149-2014 Calibration Specification for Cardiac Defibrillators”, the FLUKE IMPULSE 7000DP Defibrillator Analyzer was applied to obtain data, and then a comprehensive analysis method with graphs and table was used to analyze these test data. Results: Through analyzed the detection data of delivered energy to grasp the whole performance situation of using equipment and to enhance the cognition of technique and application level for Defibrillator Analyzer. Conclusion: Through analyzes data to find out the changing regular of delivered energy and the difference phenomenon of detection data distribution, and to elaborate the reason of phenomenon or influence factor. Through detects performance to screen out the poorest equipment of delivered energy so as to avoid medical risk that comes from hidden danger of equipment performance in the process of clinical therapy.[
出处
《中国医学装备》
2018年第1期42-46,共5页
China Medical Equipment
关键词
心脏除颤器
释放能量
数据分析
失准
除颤分析仪
Cardiac defibrillator
Delivered energy
Data analysis
Misalignment
Defibrillation analyzer