宝石能谱计算机自动智能辅助技术联合低剂量个性化扫描用于双期胸部CT增强检查的图像质量评价 被引量:6

Evaluation of image quality for chest enhanced CT scan using gemstone spectral imaging assist associated with low-dose personalized scan
下载PDF
导出
摘要 目的:评价宝石能谱计算机自动智能辅助技术(GSI assist)联合低剂量个体化扫描方案用于双期胸部CT增强扫描患者的图像质量。方法:120例行双期能谱CT增强扫描的患者分为研究组与对照组,每组60例,再分别按照体重指数(BMI)分为偏瘦、正常和超重3个亚组。患者均行美国GE宝石能谱CT扫描。研究组采用GSI assist扫描模式,对比剂碘注射剂量按300 mg/kg计算;对照组采用常规120 kVp扫描模式,碘注射剂量按450 mg/kg计算。对两组图像质量分别进行主观评分和客观评分(CNR),并记录辐射剂量。结果:按BMI分层后,研究组动脉期和静脉期低能级(70 keV)图像质量主观评分与对照组相比,差异均无统计学意义,而感兴趣区CNR均高于对照组(P<0.05);辐射剂量2组间差异均无统计学意义(P>0.05)。结论:GSI assist联合低剂量个性化扫描能够有效提高胸部CT增强扫描图像质量,减少患者总对比剂用量的同时维持稳定的辐射剂量。 Aim:To evaluate the image quality for chest enhanced CT scan using gemstone spectral imaging(GSI)assist associated with low-dose personalized scan.Methods:A total of 120 patients were allocated into control group and experimental group(60 patients respectively)and were further stratified into three sub-groups according to body mass index(BMI).Spectral enhanced CT scan was performed with GE Discovery CT in both groups.GSI assist and conventional 120 kVp scan were respectively used in experimental group and control group with iodine at 300 and 450 mg/kg.Subjective image quality score and objective image quality index(CNR)as well as radiation dose were measured and recorded.Results:After being stratified according to BMI,the subjective image quality score had no significant difference in comparison with control group(P>0.05),while CNR in region of interest was higher in experimental group(P<0.05);the radiation dose in both groups showed no significant difference(P>0.05).Conclusion:GSI assist associated with low-dose personalized scan effectively improves the image quality and reduces the dosage of contrast agent without increasing radiation dose.
作者 周悦 侯平 查开继 王睿 蒲实 王明月 董军强 高剑波 ZHOU Yue;HOU Ping;ZHA Kaiji;WANG Rui;PU Shi;WANG Mingyue;DONG Junqiang;GAO Jianbo(Department of Radiology,the First Affiliated Hospital,Zhengzhou University,Zhengzhou 450052)
出处 《郑州大学学报(医学版)》 CAS 北大核心 2020年第2期262-266,共5页 Journal of Zhengzhou University(Medical Sciences)
基金 河南省卫健委医学科技攻关计划项目(201602012)。
关键词 X线计算机体层成像 辐射剂量 体重指数 X-ray computed tomography radiation dose body mass index
  • 相关文献

参考文献7

二级参考文献62

  • 1Entrikin DW,Leipsic JA,Carr JJ.Optimization of radiation dose reduction in cardiac computed tomographic angiography [J].Cardiol Rev,2011,19:163-176. 被引量:1
  • 2Baker ME,Dong F,Primak A,et al.Contrast-to-noise ratio and low-contrast object resolution on full-and low-dose MDCT:satire versus filtered back projection in a low-contrast object phantom and in the liver[J].AJR Am J Roentgenol,2012,199:8-18. 被引量:1
  • 3Yu L,Li H,Fletcher JG,et al.Automatic selection of tube potential for radiation dose reduction in CT:a general strategy[J].Med Phys,2010,37:234-243. 被引量:1
  • 4Schindera ST,Winklehner A,Alkadhi H,et al.Effect of automatic tube voltage selection on image quality and radiationdose in abdominal CT angio-aphy of various body sizes:a phantom study[J].Clin Radiol,2013,68:79-86. 被引量:1
  • 5Kalra MK,Woisetschlager M,DahlstrOm N,et al.Radiation dose reduction with Sinogram Affirmed herative Reconstruction technique for abdominal computed tomography[J].Comput Assist Tomoqr,2012,36:339-346. 被引量:1
  • 6Behrendt FF,Schmidt B,Plumhans C.et al.Image fusion in dual energy computed tomography:effect on contrast enhancement,signal-to-noise ratio and image quality in computed tomography angiography[J].Invest Radiol,2009,44:1-6. 被引量:1
  • 7Marin D,Nelson RC,Schindera ST,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm:initial clinical experience [J].Radiology.2010,254:145-153. 被引量:1
  • 8Lee S,Yoon SW,Yoo SM,et al.Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis[J].Acta Radiologica,2011,52:1101-1106. 被引量:1
  • 9Lee KH,Lee JM,Moon SK,et al.Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT [J].Radiology,2012,265:437-447. 被引量:1
  • 10Waaijer A,Prokop M,Vehhuis BK,et al.Circle of Willis at CT angiography:dose reduction and image quality-reducing tube voltage and increasing tube current settings [J].Radiology,2007,242:832-839. 被引量:1

共引文献102

同被引文献67

引证文献6

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部