利用免洗胶片测量CT轴向和螺旋扫描的剂量分布 被引量:1

Using disposable film to measure the dose profiles in axial scanning and helical scanning with computed tomography
下载PDF
导出
摘要 目的:测量CT轴向扫描和螺旋扫描两种模式的剂量分布及其影响因素。方法:利用Gafchromic免洗胶片测量GE LightSpeed 64排CT的剂量分布。胶片经CT机扫描曝光后,利用HP scanjet 7400c扫描仪扫描获得数字图像,并利用ImageJ软件进行图像处理,以获得CT剂量分布。结果:在轴向扫描过程中,峰值半高宽与标称射束准直宽度偏差小于4mm,视野大小和过滤器形状对轴向剂量分布的影响可以忽略。在螺旋扫描过程中,过扫描范围的大小主要与射束准直宽度有关,螺距、重建层厚和视野对过扫描范围的影响较小。结论:免洗胶片可用于快速检测CT轴向和螺旋扫描的剂量分布。 Objective To measure the dose profiles in axial scanning and helical scanning with computed tomography (CT), and investigate the related influence factors. Methods Gafchromic disposable film was used to measure the dose profile of GE LightSpeed 64 CT scanner. After being exposed by CT scanner, the disposable film was scanned with HP scanjet 7400c to obtain digital image, and then ImageJ software was applied to process the digital image for obtaining the dose profile. Results In axial scanning mode, the deviation between full width at half-maximum and nominal beam collimation width was within 4 mm, and moreover, the effects of field of view and types of filters on the dose profile in axial scanning were ignorable. In helical scanning mode, the overranging was mainly dependent on beam collimation width, and moreover, pitch, reconstruction slice thickness and field of view had trivial effects on overranging. Conclusion Disposable film is capable of rapidly measuring the dose profiles in axial scanning and helical scanning with CT.
作者 戴兴 徐加利 梁保辉 DAI Xing;XU Jiali;LIANG Baohui(School of Medical Imaging,Benbu Medical College,Bengbu 233000,China;Department of Radiology,the First Affiliated Hospital of Bengbu Medical College,Bengbu 233004,China)
出处 《中国医学物理学杂志》 CSCD 2019年第5期529-533,共5页 Chinese Journal of Medical Physics
基金 安徽高校自然科学研究重点项目(KJ2017A239) 蚌埠医学院校级科研课题(byky1639)
关键词 计算机断层成像 免洗胶片 轴向扫描 螺旋扫描 computed tomography disposable film axial scanning helical scanning
  • 相关文献

参考文献1

二级参考文献24

  • 1傅征,梁铭会.数字医学概论[M].北京:人民卫生出版社,2009. 被引量:35
  • 2郑钧正.医学影像数字化所面临新课题应对刍议.世界医疗器械(IMD),2010,. 被引量:3
  • 3郑钧正.放射诊疗基本信息资源的开发与利用.世界医疗器械(IMD),2010,. 被引量:2
  • 4郑钧正 贺青华 李述唐 等.“九五”期间全国医疗照射水平调查研究.中华放射医学与防护杂志,2000,20:2-7. 被引量:13
  • 5郑钧正.x射线诊断和介入放射学的放射防护与安全[M]//郑钧正主编.电离辐射医学应用的防护与安全.北京:原子能出版社,2009. 被引量:2
  • 6Matsumoto K, Jinzaki M, tanami Y, et al. Vitual monochromatic spectral imaging with fast kilovohage switching: improved image quality as compared with that obtained with conventional 120 kVp CT[J]. Radiology, 2011, 259 (1):257 - 262. 被引量:1
  • 7UNSCEAR. UNSEAE 2008 Report Sources and effects of ionizing radiation, Vol. I [ M ]. New York : UN, 2010. 被引量:1
  • 8ICRP. ICRP Publication 87 Managing patient dose in computed tomography [ M ]. Oxford: Pergamon Press, 2001. 被引量:1
  • 9UNSCEAR. UNSEAE 2000 Report Sources and effects of ionizing radiation, Vol. I : Sources [ M ]. New York:UN, 2000. 被引量:1
  • 10郑钧正,岳保荣,李述唐,等.我国"九五"期间x射线诊断的医疗照射水平[J].中华放射医学与防护杂志,2000,20(增刊):818-S20. 被引量:1

共引文献29

同被引文献13

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部