基于DICOM文件信息估算多排螺旋CT剂量方法的探讨 被引量:4

Discussions of Multi-Detector Computed Tomography (MDCT) dose assessment based on information in DICOM images
原文传递
导出
摘要 目的 建立基于DICOM文件信息估算X射线CT剂量的方法,并验证该方法的可靠性.方法 以临床常用多排螺旋CT为研究对象,采用java语言开发DICOM图像文件信息读取软件,实现自动化批量读取在不同扫描模式下胸部、腹部X射线CT螺旋扫描的每幅图像中的剂量相关信息,结合准直、mAs、螺距等扫描参数,计算每次扫描检查的容积CT剂量指数(CTDIvol)和DLP,并与X射线CT的显示值进行比较.结果 对于胸部螺旋扫描,基于DICOM图像文件信息计算得到的CTDIvol与显示值误差为-2%~8%,DLP与显示值误差为-2%~5%;对于腹部螺旋扫描,CTDIvol计算值与显示值误差在0~ 2%,DLP与显示值误差在1% ~ 3%.计算得到的CTDIvol与显示值误差为-2%~2%,DLP与显示值误差为-2% ~3%.结论 本研究所建立的基于DICOM文件信息估算多排螺旋CT剂量的方法值得推广和应用. Objective To establish an accuracy test method for MDCT dose assessment based on information in DICOM images.Methods The type of MDCT studied in this paper was widely used in clinical practice.A software package developed by java language was used to automatically read doserelated information from DICOM files of MDCT.The CTDIvol and DLP of each pectoral or abdominal scan was calculated based on these information and the basic scan parameters such as collimation,mAs and pitch.The calculated values were compared with the displayed values.Results For pectoral scans,the difference between the calculated and displayed values was between-2%-8% for CTDIvol,and-2%-5% for DLP.For abdominal scan,the difference between the calculated and displayed values was 0-2% for CTDIvol,and-2%-3% for DLP.Conclusions This method is useful for MDCT dose assessment and is worth disseminating its application for general use.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2016年第2期143-148,共6页 Chinese Journal of Radiological Medicine and Protection
基金 国家自然科学基金,National Natural Science Foundation of China
关键词 DICOM 多排螺旋CT 剂量估算 DICOM Multi-detector computed tomography (MDCT) Dose assessment
  • 相关文献

参考文献16

  • 1International Atomic Energy Agency. Radiation protection in newer medical imaging techniques:CT colonography. IAEA safety reports series No 61[R]. Vienna:IAEA, 2008. 被引量:1
  • 2International Atomic Energy Agency. Radiation protection in newer medical imaging techniques:PET/CT. IAEA Safety reports series No 58[R]. Vienna:IAEA, 2008. 被引量:1
  • 3International Commission on Radiological Protection. Managing patient dose in computed tomography. ICRP Publication 87[R]. Oxford:Pergamon Press, 2001. 被引量:1
  • 4International Commission on Radiological Protection. Managing patient dose in multi-detector computed tomography. ICRP Publication 102[R]. Oxford:Pergamon Press, 2007. 被引量:1
  • 5National Committee on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States. NCRP Report No 160[R]. Bethesda:NCRP,2009. 被引量:1
  • 6International Atomic Energy Agency. Dosimetry in diagnostic radiology:an international code of practice. IAEA Technical reports series No 457[R]. Vienna:IAEA, 2007. 被引量:1
  • 7Tessa SC, Stefan LZ, Scott RS, et al. Informatics in radiology:RADIANCE:an automated, enterprise-wide solution for archiving and reporting CT radiation dose estimates[J]. Radio Graphics, 2011, 31(7):1833-1846. DOI:10.1148/rg.317115048. 被引量:1
  • 8Tsalafoutas A, Metallidis S. A method for calculating the dose length product from CT DICOM images[J]. Br J Radiol, 2011, 84(1004):236-243. DOI:10.1259/bjr/37630380. 被引量:1
  • 9王彬,郑钧正,高林峰,卓维海,姚杰,钱爱君.上海市医用X射线CT的应用频率及其分布研究[J].辐射防护,2013,33(2):65-73. 被引量:30
  • 10高林峰,郑钧正,卓维海,王彬,曲良勇,姚杰,钱爱君.医用X射线CT主要检查类型所致成年受检者剂量研究[J].辐射防护,2013,33(2):74-81. 被引量:24

二级参考文献53

  • 1傅征,梁铭会.数字医学概论[M].北京:人民卫生出版社,2009. 被引量:35
  • 2郑钧正.医学影像数字化所面临新课题应对刍议.世界医疗器械(IMD),2010,. 被引量:3
  • 3郑钧正.放射诊疗基本信息资源的开发与利用.世界医疗器械(IMD),2010,. 被引量:2
  • 4郑钧正 贺青华 李述唐 等.“九五”期间全国医疗照射水平调查研究.中华放射医学与防护杂志,2000,20:2-7. 被引量:13
  • 5郑钧正.x射线诊断和介入放射学的放射防护与安全[M]//郑钧正主编.电离辐射医学应用的防护与安全.北京:原子能出版社,2009. 被引量:2
  • 6Matsumoto K, Jinzaki M, tanami Y, et al. Vitual monochromatic spectral imaging with fast kilovohage switching: improved image quality as compared with that obtained with conventional 120 kVp CT[J]. Radiology, 2011, 259 (1):257 - 262. 被引量:1
  • 7UNSCEAR. UNSEAE 2008 Report Sources and effects of ionizing radiation, Vol. I [ M ]. New York : UN, 2010. 被引量:1
  • 8ICRP. ICRP Publication 87 Managing patient dose in computed tomography [ M ]. Oxford: Pergamon Press, 2001. 被引量:1
  • 9UNSCEAR. UNSEAE 2000 Report Sources and effects of ionizing radiation, Vol. I : Sources [ M ]. New York:UN, 2000. 被引量:1
  • 10郑钧正,岳保荣,李述唐,等.我国"九五"期间x射线诊断的医疗照射水平[J].中华放射医学与防护杂志,2000,20(增刊):818-S20. 被引量:1

共引文献46

同被引文献35

引证文献4

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部