摘要
目的通过查看日立7180生化仪报警提示,对出现的假结果进行分析和处理,得出正确的检测结果。方法对于测定中出现线性报警提示的检测结果,通过稀释以及重新预处理标本来测定血清脂肪酶值。结果通过观察反应曲线和各监测点吸光度值的变化对仪器报警显示进行分析,直至测定出准确无报警提示结果。结论全自动生化仪实际使用过程中会出现各种报警,每种报警都可能直接影响检测结果,需工作人员认真对待,直至排除所有因素,得出无报警结果,为临床诊疗提供更准确和有价值的检测结果。
Objective By viewing the Hitachi7180biochemical instrument alarm,analysis and processing of false results,get the correct results.Methods The serum lipase value was determined by diluting and re treating the sample for the detection results of linear warning prompt in the determination.Results Observing reaction curve and the variation of absorbance value of every monitoring points until the right,alarm-free results are concluded.Conclusion There will be many kinds of alarms in the actual use of biochemical analyzer.Each alarm may directly affect the test,which shall be taken seriously by the operator until all possible factors are excluded,until all factors are eliminated,no alarm results detected,to provide more accurate and valuable test results for clinical diagnosis and treatment.
作者
万俊梅
祝丹丹
罗琴梅
WAN Junmei;ZHU Dandan;LUO Qinmei(Department of Clinical Laboratory,People′s Hospital of Shayang County,Shayang,Hubei 448200,China)
出处
《国际检验医学杂志》
CAS
2017年第A02期117-119,共3页
International Journal of Laboratory Medicine