期刊文献+

工业原子力显微镜快速进针系统的设计

Design of Industrial AFM Fast Tip Approach System
下载PDF
导出
摘要 摘要为了满足工业原子力显微镜高效检测的要求,提出了一种粗细相结合的自动进针方法。设计了快速的粗进针和慢速的精细进针二级进针控制系统硬件结构。针对工业半导体样品的特点,通过预标定方法保证粗进针动作快速到位。在精细进针阶段采用升降台步进动作和实时闭环检测,当检测到激光信号变化的值超过setpoint值时,完成进针操作。通过实验测试表明该方法有效减少了进针时间,提高了原子力显微镜的进针效率。 In order to meet the industrial requirements for the eff^cient test of atomic force microscope, a fast tip approach method composed of rough approach and fine approach method is introduced. The hardware controlling system for the two level approaching including rough and fine section wasdesigned. According to the characteristics of the semiconductor industry sample, rough approach movement wasdone by the pre-calibration method. Stage step moving and close-loop real-time detection were acted during the fine tip approaching.When the laser signal value exceeds the value of setpoint, the approaching process is finished. Experiments showed that this method is effective to reduce the approaching time and to improve the efficiency.
出处 《现代科学仪器》 2016年第3期29-33,共5页 Modern Scientific Instruments
基金 国家面上基金支持项目:51375470 863计划项目:2012AA041204
关键词 原子力显微镜 进针控制 激光检测 PC-104控制器 AFM Approaching control Laser detection PC-104 controller
  • 相关文献

参考文献8

二级参考文献54

  • 1朱杰,孙润广.原子力显微镜的基本原理及其方法学研究[J].生命科学仪器,2005,3(1):22-26. 被引量:65
  • 2白春礼.扫描隧道显微术及其应用[M].上海:上海科学技术出版社,1991.1-14. 被引量:6
  • 3Binnig G ,Quate C F,Gerber C. Atomic force microscope[J]. Physical review letters, 1986, 56(9). : 930 -933. 被引量:1
  • 4Albrecht T R,Grutter P, Home D, et al. Frequencymodulation detection using high-Q cantilevers forenhanced force microscope sensitivity [ J]. Journal ofApplied Physics, 1991 , 69(2). : 668 -673. 被引量:1
  • 5Zhong Q,Inniss D,Kjoller K, et al. Fracturedpolymer/silica fiber surface studied by tapping modeatomic force microscopy [ J]. Surface Science Letters,1993, 290( 1). ; L688 -L692. 被引量:1
  • 6Overney R M, Meyer E, Frommer J, et al. Forcemicroscopy study of friction and elastic compliance ofphase-separated organic thin films [J]. Langmuir,1994,10(4). : 1281 -1286. 被引量:1
  • 7Martin Y, Williams C C, Wickramasinghe H K. Atomicforce microscope-force mapping and profiling on a sub100 - A scale [ J]. Journal of Applied Physics, 1987 ,61(10). ; 4723 -4729. 被引量:1
  • 8Garcfa R,Perez R. Dynamic atomic force microscopymethods[ J]. Surface Science Reports, 2002,47 (6).:197 -301. 被引量:1
  • 9Hansma P K, Cleveland J P, Radmacher M,et al.Tapping mode atomic force microscopy in liquids [ J].Applied Physics Letters, 1994, 64(13). : 1738 1740. 被引量:1
  • 10Magonov S N, Elings V,Whangbo M H. Phase imagingand stiffness in tapping-mode atomic force microscopy[J]. Surface Science, 1997, 375(2). : L385 -L391. 被引量:1

共引文献31