摘要
目的利用自行搭建的超长扫描深度的频域光学相干断层扫描(SD-OCT)活体测量小鼠眼球的眼轴长度、曲率等各眼内生物学参数,并评估其重复性和再现性。方法实验研究。本研究自主研发的超长扫描深度SD—OCT扫描深度7.3mm(空气中),系统轴向扫描(A—Scan)速度28kHz,轴向分辨率7.5um,横向分辨率19um。共有5只40日龄的C57BL/6小鼠(10眼)经历2天次的测量。第l天,由操作者1先执行测量步骤,连续拍摄2幅图后,由操作者2紧接着拍摄1幅图。在3次拍摄中,每次都保证前后拍摄的时间耽搁小于5min。2d以后.所有的小鼠再由操作者1拍摄1幅图,随后每只小鼠再进行1次时域OCT(TD—OCT)拍摄。所有测量都在下午2:00到4:00之间进行。测量的重复性和再现性评估都采用重复性系数、组内相关系数ICCs、Wilcoxon配对测试及绘制Bland.Altman图来综合分析。SD—OCT和TD.OCT的测量结果比较采用独立样本t检验。结果用SD—OCT成功获得了清晰、高质量的小鼠跟球一次性成像。除角膜、前房外,其他数据接近于TD—OCT。前房、晶状体、玻璃体和眼轴的重复性/再现性系数均小于3%且对应的组内相关系数(ICCs)几乎都大于0.75,角膜和视网膜的重复性/再现性系数小于5%且对应的ICCs相对较小。除操作者l第1天获取的2组角膜数据之间差异有统计学意义外(压一2-310,P〈0.05),其余差异均无统计学意义。眼轴等生物学参数的Bland.Ahman分析图显示所有的数据点都分布在95%可信区间内。各个曲率的重复性,再现性系数均小于5%。角膜前表面曲率半径和晶状体前、后表面曲率半径的ICCs基本大于0.75,而角膜后表面曲率半径的ICC大于0.45。Bland—Altman图显示曲率再现性的数据点约有5%分布在95%可信区间外。结论利用SD—OCT,可获得高清晰的从小鼠角膜到视网�
Objective To assess the accuracy, precision, repeatability, and reproducibility of biometric measurement of the mouse eye made by extended scan depth spectral domain optical coherence tomography (SD-OCT). Methods It was an experimental study. An SD-OCT with an extended depth range of 7.3 mm, A-scan speed of 28 kHz, axial resolution of 9.0 um and transversally 19 um was developed. Five 40-day-old C57BL/6 mice (10 eyes) were imaged with a custom built, extended scan depth SD-OCT in 2 sessions. Two scans were performed by operator 1 and a following scan by operator 2 at the first session, to determine repeatability and interoperator reproducibility. Intersession reproducibility was assessed by an additional scan performed by operator 1 2 days later. The data were compared to those from time-domain OCT (TD-OCT) by independent t test. All experiments were performed between 2:00 PM and 4:00 PM. The repeatability and reproducibility were evaluated by coefficient of repeatability/reproducibility, intraclass correlation coefficients (ICCs), Wilcoxon matched-pairs test and Bland-Altman graph. Results High-quality OCT images of the entire mouse eyes with high resolution were acquired. The ocular dimensions from TD-OCT measurements were close to these obtained by SD-OCT. The overall coeffieients of repeatability/reproducibility for anterior chamber depth, lens thickness, vitreous chamber and axial length were all less than 3% and relevant intraclass correlation coefficients (ICCs) were nearly greater than 0.75, whereas these were less than 5% and ICCs were smaller for corneal thickness and retinal thickness. Wileoxon analysis showed there was no statistically significant difference between any 2 groups for all parametres except corneal thickness of repeatability (Z=-2.310, P〈0.05). From the graph of differences against mean for the axial lengths, it can be seen that 100% of the values fall within 2 SDs of the mean. The overall coefficients of repeatability/reproducibility for curvatures were les
出处
《中华眼视光学与视觉科学杂志》
CAS
2013年第6期347-353,共7页
Chinese Journal Of Optometry Ophthalmology And Visual Science
关键词
体层摄影术
光学相干
生物学测量
曲率
小鼠
Tomography,optical coherence
Biometric measurement
Curvature
Mice