摘要
本文介绍了智能化呼吸机近4年在临床应用55577天中存在1125次安全隐患的处理方法及预防措施。统计出近4年呼吸机的平均使用效率达67.83%,并介绍了呼吸机使用前的"九步法"检查法,经临床验证,可有效降低故障率,有较高的临床推广应用价值。
This paper introduces the processing method and prophylactic intervention of 1,125 potential safety hazards in 55,577 days clinical application of smart ventilator in latest 4 years.It is counted that the average service efficiency is 67.83%.The Operation Verification Procedure before using ventilator is describded,it can decrease the failure rate effectively and has vital clinical value by clinical verification.
出处
《中国医疗设备》
2011年第3期97-99,共3页
China Medical Devices
关键词
呼吸机
临床应用
预防措施
ventilator
clinical application
preventive intervention