期刊文献+

氟-肝素表面修饰人工晶状体的制备及物理光学性能的研究 被引量:4

Preparation of F heparin surface modified intraocular lens and study onits physical and optical properties
下载PDF
导出
摘要 目的以氟-肝素对聚甲基丙烯酸甲酯(PMMA)人工晶状体进行表面修饰,并测定其物理光学性能。方法利用载能离子束联合低温等离子体技术对PMMA人工晶状体进行氟-肝素表面修饰,通过物理光学方法对不同剂量表面修饰的人工晶状体进行透光率、分辨率、屈光度及袢的抗拉强度的测定。结果氟-肝素修饰人工晶状体氟离子剂量为2×1014~5×1014ions/cm2,其机械、光学性能和未修饰人工晶状体比较无明显区别,在短波光线及紫外线区(360~490nm)修饰人工晶状体的透光率明显低于未修饰人工晶状体。结论氟-肝素修饰人工晶状体有效地阻挡了短波光线及紫外线,起到保护视网膜的作用。 ObjectiveTo prepare the surface modified IOL with F ions and heparin and to study the physical and optical properties.MethodsUsing the technique of carrier energy ion beam combined low temperature and low pressure plasma to modify the surface of PMMA IOL with F ions and heparin.The transmittance,resolution,diopter and anti compression intensity of loops were determined by using physical and optical methods.ResultsThe surface modified IOLs with F ions 2×10 14 ~5×10 14 ions/cm 2 and heparin showed no difference from the control PMMA IOLs in their mechanical and optical properties.However,the transmittance of the surface modified IOLs in the short wave light zone and ultraviolet rays(360~490nm) was obviously lower than the unmodified IOLs.ConclusionF-heparin modified IOL can decrease the harmful effect of the short wave light and ultravoilet rays to the retina.
出处 《眼科研究》 CSCD 1999年第2期95-97,共3页 Chinese Ophthalmic Research
关键词 人工晶状体 表面修饰 白内障 人工晶体植入术 ntraocular lens surface modification physical and optical property determination