两种电子根尖定位仪准确性的比较 被引量:1

A Comparison of the accuracy of working length determination between two electronic apex locators
下载PDF
导出
摘要 目的比较临床常用的Neosono Co-pilot和JustyⅡ两种电子根尖定位仪测定根管工作长度的准确性。方法选取临床上拟行根管治疗的患牙182颗,共397条根管,先后用Neosono Co-pilot和JustyⅡ根尖定位仪测定根管长度,根据后一种方法测得的长度插扩大针拍摄定位X线片,测量扩大针尖至根尖的距离。结果结果Neoso-no Co-pilot距根尖1 mm内的准确率为93.2%,JustyⅡ的准确率为94.5%,差异无统计学意义(P>0.05);两种定位仪对不同疾病患牙的准确性差异无统计学意义。结论两种电子根尖定位仪用作测量根管工作长度准确、简便而实用。 Objective To compare the accuracy of working length determination using two electronic apex locators: Neosono Co-pilot and Justy 11. Methods 397 canals in 182 teeth to do root canal therapy were elected. Working length determination was performed using two apex locators: Neosono Co-pilot and Justy 11. The actual tooth length was calculated by X-ray with a file which length was counted by the later locator. To evaluate the accuracy of both electronic apex locators, each electroni- cally determined distance was compared with the actual tooth length calclated by X-ray, and the data analyzed using the SPSS 12. 0 for windows and the chi-square test. Results The accuracy of Neosono Co-pilot was 93.2% and Justy 11 was 94. 5%. There was no statistically significant difference between two electronic apex locators and there was no statistically significant difference among working length determination of different diseases using two electronic apex locators. Conclusion The practice of using these two electronic apex locators in determination of working length is useful and reliable.
出处 《广州医药》 2007年第6期20-22,共3页 Guangzhou Medical Journal
关键词 电子根尖定位仪 根管工作长度 根管 准确性 Tooth Electronic Apex locator Working length Canal Accuracy
  • 相关文献

参考文献11

二级参考文献21

  • 1覃荣林,朱霞.影响根管长度电测仪测量效果的因素分析[J].上海生物医学工程,1996,17(3):27-28. 被引量:14
  • 2张举之.口腔内科学(第3版)[M].北京:人民卫生出版社,1995.390-391. 被引量:124
  • 3凌均qi.牙髓病学[M].北京:人民卫生出版社,1998.43. 被引量:5
  • 4[1]Huang L.An experimental study of the principle of electronic root canal measurement.J Endod,1987,13:60-4. 被引量:1
  • 5[2]Hülsmann M,Pieper K.Use of an electronic apex locator in the treatment of teeth with incomplete root formation.Endod Dent Traumatol,1989,5:238-41. 被引量:1
  • 6[3]Frank AL,Torabinejad M.An in vivo evaluation of Endex electronic apex locator.J Endod,1993,19:177-9. 被引量:1
  • 7[4]Mayeda DL,Simon JHS,Aimer DF,et al.In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator.J Endod,1993,19:545-8. 被引量:1
  • 8[5]Kobayashi C,Suda H.New electronic canal measuring device based on the rat iomethod.J Endod,1994,20:111-4. 被引量:1
  • 9[6]Fernando G,Ana C,Susana M,et al.J Endod,2002,28,461-3. 被引量:1
  • 10[7]Dunlap CA,Remeikis NA,BeGole EA,et al.An in vivo evaluation of an electro nicapex locator that uses the ratio method in vital and necrotic canals.J Endod,1998,24:48-50. 被引量:1