三种根管长度测量仪测定根尖吸收牙齿工作长度准确性的比较研究 被引量:11

In vitro measurement accuracy of electronic apex locator in teeth with apical root resorption
下载PDF
导出
摘要 目的 比较三种不同根管长度测量仪测定根尖区吸收牙齿工作长度结果的准确性。方法 将 6 0颗离体前牙平均分成三组 ,用 70 0 # 裂钻深入根尖孔钻入 3mm ,破坏根尖狭窄处。将三组牙齿分别放置于测量槽中 ,使用RootZX、Neosonic和JustyⅡ测量其工作长度。结果 所得出的数据与肉眼下准确测量所得出的数据比较 ,A组在± 0 .5mm 6 5 .0 %、± 1mm 85 .0 %、± 1.5mm 10 0 .0 %。B组在± 0 .5mm 6 0 .0 %、± 1mm 85 .0 %、± 1.5mm10 0 .0 %。C组在± 0 .5mm 75 .0 %、± 1mm 90 .0 %、± 1.5mm 10 0 .0 %。结论 三种测量仪测量值具有准确性 ,A组。 Objective This study was to evaluate the accuracy of three kinds electronic apex locators to determine the working length in teeth with simulated apical root resorption.Methods Sixty anterior teeth with mature apices were divided into three groups. An irregular cavity defect was drilled at the apies of each tooth simulating an apical root resorption.Using the Root ZX (group A),Neosonic (group B),and JustyⅡ(group C),the working lengths was measured,comparing the electronic readings with direct visual measurements.Results The Root ZX was 65.0% , 85.0% and 100.0% accurately to within 0.5mm ,1mm,and 1.5mm of the direct visual measurements.The Neosonic was 60.0% , 85.0% and 100.0% .The JustyⅡwas 75.0% ? 90.0% ? 100.0% .Conclusion All 3 electronic apex locators were useful for working length measurement in teeth with apex resorption.Statistically,no significant differences were observed among group A,group B,and group C.
出处 《现代口腔医学杂志》 CAS CSCD 2004年第2期130-131,共2页 Journal of Modern Stomatology
关键词 根管长度测量仪 测定 根尖吸收 牙齿工作长度 准确性 根尖狭窄 Electronic apex locator Apical constriction Working lengths
  • 相关文献

参考文献8

  • 1[1]Huang L.An experimental study of the principle of electronic root canal measurement.J Endod,1987,13:60-4. 被引量:1
  • 2[2]Hülsmann M,Pieper K.Use of an electronic apex locator in the treatment of teeth with incomplete root formation.Endod Dent Traumatol,1989,5:238-41. 被引量:1
  • 3[3]Frank AL,Torabinejad M.An in vivo evaluation of Endex electronic apex locator.J Endod,1993,19:177-9. 被引量:1
  • 4[4]Mayeda DL,Simon JHS,Aimer DF,et al.In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator.J Endod,1993,19:545-8. 被引量:1
  • 5[5]Kobayashi C,Suda H.New electronic canal measuring device based on the rat iomethod.J Endod,1994,20:111-4. 被引量:1
  • 6[6]Fernando G,Ana C,Susana M,et al.J Endod,2002,28,461-3. 被引量:1
  • 7[7]Dunlap CA,Remeikis NA,BeGole EA,et al.An in vivo evaluation of an electro nicapex locator that uses the ratio method in vital and necrotic canals.J Endod,1998,24:48-50. 被引量:1
  • 8[8]Shabahang S,Goon WWY,Gluskin AH.An in vivo evaluation of Root ZX electronic apex locator.J Endod,1996,22:616-8. 被引量:1

同被引文献58

引证文献11

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部