期刊文献+

新一代冲击加速度国家基准装置的研究与建立 被引量:17

Establishment of new generation national primary standard device for shock acceleration calibration
下载PDF
导出
摘要 研究建立的新一代冲击加速度国家基准装置,在冲击加速度峰值20~1 000 000 m/s2和脉冲持续时间20~200μs范围内,实现了加速度计冲击灵敏度激光干涉法的高精度校准。介绍了装置的结构组成和解决的关键技术问题,描述了中小冲击装置加速度峰值和脉冲持续时间的控制功能,给出了装置的主要技术指标和实验数据。新一代冲击加速度基准装置可以解决国民经济各领域对冲击计量提出的宽量程、高准确度的量值溯源需求问题。 The new generation national primary standard device of shock acceleration was studied and established. It can realize high precision calibration of the sensitivity of shock accelerometer by laser interferometry in the acceleration peak range from 20 m/s2 to 1000000 m/s2 and the pulse-duration range from 20 μs to 200 μs. The device structure and the key technology to be solved were introduced. The performance of acceleration-peak and pulse-duration control for small-medium shock devices were described. The main specifications and experimental results were provided. The new generation national primary standard device can meet the demand of shock quantity traceability in wide range and high accuracy in all areas of national economy.
出处 《振动与冲击》 EI CSCD 北大核心 2013年第10期124-129,共6页 Journal of Vibration and Shock
基金 质检公益性行业科研专项项目(200810007)
关键词 计量学 冲击加速度计量基准 冲击加速度灵敏度 加速度峰值 冲击脉宽持续时间 外差正交激光干涉仪 测量不确定度 Accelerometers Calibration Laser interferometry Measurements Uncertainty analysis
  • 相关文献

参考文献10

  • 1Link A,von Martens H J.Accelerometer Identification Using ShockExcitation[].Measeurement.2004 被引量:1
  • 2Methods for the calibration of vibration andshock transducers-part 13:primary shock calibration by laserinterferometry. ISO 16063-13 . 2001 被引量:1
  • 3胡红波,于梅.基于高冲击激励的加速度计参数辨识的研究[J].传感技术学报,2012,25(4):487-491. 被引量:8
  • 4LINK A,T (A|¨)UBNER A,WABINSKI W,et al.Modelling accelerometers for transient signals using calibration measurements upon sinusoidal excitation[].Measurement.2007 被引量:1
  • 5Elster C,Link A,Bruns T.Analysis of dynamic measurements and determination of time-dependent measurement uncertainty using a second-order model[].Measurement Science and Technology.2007 被引量:1
  • 6Link A,von Martens H J,Wabinski W.New method for absolute shock calibration of accelerometers[].Proceedings of SPIE the International Society for Optical Engineering.1998 被引量:1
  • 7Link A,von Martens H J.Calibration of accelerometers byshock excitation and laser interferometry[].Shock and Vibration.2000 被引量:1
  • 8Bruns T,Link A,Elster C.Current Developments in theField of Shock Calibration[].Procth IMEKO WorldCongress.2006 被引量:1
  • 9Link A,Tubner A,Wabinski W,et al.Calibration ofaccelerometers:determination of amplitude and phaseresponse upon shock excitation[].Measurement Science and Technology.2006 被引量:1
  • 10胡红波,于梅.Hopkinson杆冲击激励系统的原理与应用[J].计量技术,2011(2):36-38. 被引量:6

二级参考文献17

  • 1徐科军,陶有军,邹锦.机器人腕力传感器辩识建模与动态补偿[J].计量学报,1994,15(2):138-142. 被引量:13
  • 2裴纺霞,王恒辉.基于数据融合的加速度传感器的静态模型辨识[J].传感技术学报,2005,18(4):894-896. 被引量:2
  • 3ISO16063-22:2005 Methods for calibration of vibration and shock transducers-part 22:Shock calibration by comparison to a reference transducer[S]. 被引量:1
  • 4ISO 2001 International Standard 16063-13 Methods for the Calibration of Vibration and Shock Transducers--Part 13: Primary Shock Calibration Using Laser Interferometry [ S ]. (C, eneva:Internatianal Organization for Standardization),2001. 被引量:1
  • 5ISO 1999 International Standard 16063-11 Methods for the Calibration of Vibralion and Shock Transducers--Part 11: Primary Vibration Calibration by Laser Interferometry [ S ]. (Geneva: International Organization for Standardization), 1999. 被引量:1
  • 6Hessling J P. Dynamic Metrology--An Approach to Dynamic Evaluation of Linear Time-Invariant Measurement Systems [ J ]. Measurement Sci & Tech ,2008,19 : 1-7. 被引量:1
  • 7Link A,von Martens H J. Accelerometer Identification Using Shock Excitation [ J ]. Measeurement,2004,35 : 191 - 199. 被引量:1
  • 8Thomas Bruns, Afred Link, Franko Schmahling, et al. Calibration of Aceelerometers Using Parameter Identification--Targeting a Versatile New Standard [ C ]//XIX IMERO World Congress, Fundamental and Applied Metrology, September 6 - 11, Lisbon, Portugal, 2009. 被引量:1
  • 9Bruns T,Link A, Elster C. Current Developments in the Field of Shock Calibration [ C ]//Proc. 18th IMEKO World Congress ( Rio de Janeiro, Brazil) ,2006. 被引量:1
  • 10Ueda K, Umeda A. Dynamic Response of Shock Accelerometer Measured Using Davies Bar Technique and Laser Interferometry [ C]//Pro. Seventh Int. Cong. on Exp, Mech. Las Vegas USA, 8- 11 June 1992 : 1666-1673. 被引量:1

共引文献11