摘要
西门子Force双源CT属于高端CT,设备整体配置高、性能佳、应用范围广、稳定性好,在医院运行中发挥了巨大作用。由于设备使用量大,故障时有发生,且设备集成程度较高、整体维修难度大。本文详细分析了使用过程中发生的水冷系统、患者检查床、探测器电源3例故障的故障排除过程,并通过更换三通阀、维护升降电机轴承、焊接X10接头解决故障,节省了维修费用,并为同行维修此类故障提供参考。
Siemens Force dual source CT is a high-end CT,with high overall configuration,good performance,wide application range,and good stability,playing a huge role in the operation of hospital.Due to the high usage of equipment,failures have occurred from time to time,and the high level of equipment integration makes overall maintenance difficult.This paper analyzed in detail the troubleshooting process of three cases of water cooling system,patient inspection bed and detector power supply faults occurred during use,and solved the faults by replacing the three-way valve,maintaining the lifting motor bearing and welding X10 joint,which can save maintenance costs and provide reference for peers to repair such faults.
作者
孙小磊
肖吴进
汪缨
张晖
SUN Xiaolei;XIAO Wujin;WANG Ying;ZHANG Hui(Department of Clinical Medical Engineering,The First Affiliated Hospital with Nanjing Medical University,Nanjing Jiangsu 210029,China)
出处
《中国医疗设备》
2024年第8期177-180,共4页
China Medical Devices
关键词
双源CT
水冷系统
检查床
探测器电源
dual source CT
water cooling system
patient table
detector power supply