摘要 简要介绍了CT稳定运行的重要性,详细分析了西门子Definition双源CT的滤过器运动电动机过流罕见故障并给出了具体的排除方法,保障了机器的正常运行,节省了维修成本,为同行维修类似故障提供了参考。 The importance of stable operation of CT machine was introduced briefly.A rare failure of filter motion motor overcurrent was analyzed in detail for Siemens Definition dual source CT,and some treatment method and feasible solution were put forward accordingly.The stability of CT machine was ensured,and the maintenance cost was saved.References were provided for the treatment of similar failures.