摘要
为了实现人体心血管多参数检测,设计了一种基于光电容积脉搏波描记法(photoplethysmography,PPG)的心血管多参数检测系统。该系统采用S5PV210处理器为控制核心,设计了指端脉搏波采集及滤波电路,提取指端脉搏波信号特征,测量人体心率、血氧饱和度、血压值。结果表明:与标准仪器对比,心率测量最大误差为5次/min;血氧饱和度测量最大误差为2%;血压测量满足ANSI/AA-MI国际电子血压仪标准。
In order to detect the human cardiovascular multi-parameter,a cardiovascular multi-parameter detection system based on photoplethysmography(PPG)was designed.In the system,S5PV210 processor was used as the control core,and a finger-pulse wave acquisition and filter circuit were designed.Extracted the signal characteristics of finger-pulse wave to measure human heart rate,blood oxygen saturation and blood pressure.The results show that comparing with standard instruments,the maximum error of heart rate measurement is 5 beats/min,the maximum error of blood oxygen saturation measurement is 2%.Blood pressure measurement meets ANSI/AA-MI international electronic sphygmomanometer standard.
作者
张振强
陈真诚
顾爽
韩慧茹
ZHANG Zhen-qiang;CHEN Zhen-cheng;GU Shuang;HAN Hui-ru(School of Electronic Engineering and Automation,Guilin University of Electronic Technology,Guilin 541004,China)
出处
《科学技术与工程》
北大核心
2021年第20期8383-8388,共6页
Science Technology and Engineering
基金
国家自然科学基金(61271119)
国家重大科研仪器研制项目(61627807)
国家科技支撑项目(2013BAI03B01)
广西创新驱动重大项目(2019AA12005)。
关键词
心血管
多参数
光电容积脉搏波描记法
指端脉搏波
信号特征
cardiovascular
multi-parameter
photoplethysmography
fingertip pulse wave
signal characteristics