摘要 为了实现3.0T颅脑磁共振(MRI)的研发突破,对颅脑MRI进行了分析。根据分析结果对主磁体线圈进行了优化,并建立数学模型,进行优化计算。根据计算结果搭建物理模型,进行电磁仿真分析。结果表明,磁场均匀度符合设计要求。因此,提出的主磁体线圈优化设计在结构减小的情况下,具有较高的磁场均匀度。 In order to achieve a breakthrough in the development of 3.0T cranial magnetic resonance imaging(MRI),this article analyzed the cranial brain MRI.According to the analysis results,the main magnet coil was optimized,and a mathematical model was established to perform optimized calculations.A physical model was built based on the calculation results and the electromagnetic simulation analysis was conducted.The analytic results show that the uniformity of the magnetic field meets the design requirements.Therefore,the proposed optimized design of the main magnet coil has a higher magnetic field uniformity under the condition of reduced structure.
作者 朱旭帅 刘志珍 刘振友 魏小钊 冯国文 Zhu Xushuai;Liu Zhizhen;Liu Zhenyou;Wei Xiaozhao;Feng Guowen(Electrical Engineering College, Shandong University, Jinan Shandong 250061,China)
出处 《电气自动化》 2021年第3期83-85,88,共4页 Electrical Automation
基金 山东省重大科技创新工程项目(2019TSLH0410)。
关键词 3.0T颅脑磁共振(MRI) 主磁体线圈 磁场均匀度 3.0T cranial magnetic resonance imaging(MRI) main magnet coil magnetic field uniformity