期刊文献+

通过巡检、电气安全检测、设备性能检测对麻醉机进行质量控制的探究 被引量:9

Research on Quality Control of Anesthesia Machine Through Inspection,Electrical Safety Testing and Equipment Performance Testing
下载PDF
导出
摘要 目的保证麻醉机的附件完整,运行质量可靠,参数监测精准,保障患者和使用者安全。方法使用FLUKE公司制造的电气安全分析仪FLUKE ESA620,气体流量分析仪FLUKE VT305,联系日常麻醉机使用过程中常见故障,从巡检、电气安全、性能参数三方面对麻醉机进行质量控制,并分析结果。结果巡检确保设备外观附件功能正常,简易处置后合格率100%;电气安全检测合格率64.3%,不合格项目为接地电阻过大;性能检测结果100%处于可接受范围内(15%),确保了设备性能有效性。结论通过对设备进行巡检,电气安全检测及设备性能检测,有效保证了麻醉机临床应用质量。 Objective Ensure that the accessories of the anesthesia Machine are complete,the operation quality is reliable,the parameters are monitored accurately,and guarantee the safety of patients and users.Methods Using the electric safety analyzer FLUKE ESA620 and the gas flow analyzer FLUKE VT305 manufactured by FLUKE Company,the quality control of the anesthesia machine was carried out from three aspects of inspection,electric safety testing and performance parameters in connection with the common faults in the use of the daily anesthesia machine,and the results were analyzed.Results Inspection ensured that the appearance accessories of the equipment was normal,and the pass rate was 100%after simple disposal;The electrical safety test rate was 64.3%,and the unqualified item of which was grounding resistance exceeded the limits.The result of functional parameter deviation was 100%within the acceptable range(15%),ensuring that the equipment functional parameters were valid.Conclusion Through the inspection,electrical safety testing and equipment performance testing of the equipment,the clinical application quality of the anesthesia machine is effectively guaranteed.
作者 谢杲 XIE Gao(Department of Clinical Medical Engineering,Ningbo Yinzhou No.2 Hospital,Ningbo Zhejiang 315100,China)
出处 《中国医疗设备》 2021年第3期62-65,共4页 China Medical Devices
关键词 麻醉机 质量控制 巡检 电气安全 性能检测 anesthesia machine quality control inspection electrical safety performance testing
  • 相关文献

参考文献12

二级参考文献83

共引文献66

同被引文献81

引证文献9

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部