低辐射剂量下FBP、ASIR和ASIR-V 3种不同重建算法对上腹部CT图像质量的影响 被引量:10

Effect of three different reconstruction algorithms(FBP,ASIR and ASIR-V)on the upper abdominal CT image quality with low radiation dose
下载PDF
导出
摘要 目的:评价低辐射剂量时,滤波反投影(FBP)、自适应统计迭代重建(ASIR)和新一代自适应统计迭代重建(ASIR-V)算法对上腹部CT图像质量的影响。方法:选择25例拟诊胸部病变需行CT检查患者,使用GE Discovery CT 750 HD CT机行胸部低计量扫描。所得图像行FBP、ASIR、ASIR-V后测量肝脏、脾脏、竖脊肌的CT值、背景噪声(SD),计算SNR和CNR,并行相关统计学分析。由2位医师对图像进行主观图像评价。结果:2位医师对图像的主观评价具有很好的一致性。50%ASIR-V组肝脏、脾脏和竖脊肌的SD值均低于40%ASIR和FBP(均P<0.05)。50%ASIR-V组肝脏、脾脏和竖脊肌的SNR值和CNR值均高于40%ASIR和FBP(均P<0.05)。结论:在腹部低剂量CT扫描中,50%ASIR-V与40%ASIR及FBP方法相比,降低了图像噪声,提高了图像质量。 Objective:To evaluate the effect of three different algorithms,the filtered back projection(FBP),the adaptive statistical iterative reconstruction(ASIR)and the new adaptive statistical iterative reconstruction(ASIR-V)on low radiation dose upper abdominal CT image quality.Method:25 patients underwent upper abdominal plain scans on GE Discovery CT 750 HD were included.Statistical parameters were used to analyze the evaluation parameters CT values,SD,CNR and SNR.Result:The subjective scores of the three algorithms had good consistency between reviewers.The SD values of the liver,spleen and erector spinae with 50%ASIR-V were lower than those with 40%ASIR and FBP(all P<0.05).With 50%ASIR-V,the SNR and CNR values of liver,spleen and erector spinae were statistically higher than those of 40%ASIR and FBP(all P<0.05).Conclusion:In low radiation dose abdominal CT,50%ASIR-V reduces image noise and improves image quality when compared with 40%ASIR and FBP reconstruction methods.
作者 张喜荣 贺太平 贾永军 陈静 马光明 郭长义 ZHANG Xirong;HE Taiping;JIA Yongjun;CHEN Jing;MA Guangming;GUO Changyi(Division of CT,Second Affiliated Hospital of Shaanxi University of Chinese Medcine,Xi’an 712000,China)
出处 《中国中西医结合影像学杂志》 2020年第3期305-308,共4页 Chinese Imaging Journal of Integrated Traditional and Western Medicine
基金 陕西中医药大学创新团队资助项目(2019-YS04)。
关键词 迭代重建 图像质量 体层摄影术 X线计算机 Iterative reconstruction Image quality Tomography X-ray computed
  • 相关文献

参考文献4

二级参考文献27

  • 1Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radia-tion dose optimization.Radiology,2004,230(3):619-628. 被引量:1
  • 2Heyer CM,Mohr PS,Lemburg SP,et al.Image quality and ra-diation exposure at pulmonary CT angiography with 100-or 120-kVp protocol:Prospective randomized study.Radiology,2007,245(2):577-583. 被引量:1
  • 3Kalra MK,Wittram C,Maher MM,et al.Can noise reductionfilters improve low-radiation-dose chest CT images?Pilot study.Radiology,2003,228(1):257-264. 被引量:1
  • 4Marin D,Nelson RC,Schindera ST,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:Improved imagequality and decreased radiation dose with adaptive statistical itera-tive reconstruction algorithm--initial clinical experience.Radiolo-gy,2010,254(1):145-153. 被引量:1
  • 5Pan X,Sidky EY,Vannier M.Why do commercial CT scannersstill employ traditional,filtered back-projection for image recon-struction?Inverse Probl,2009,25(12):1230009. 被引量:1
  • 6Knesaurek K,Machac J,Vallabhajosula S,et al.A new iterativereconstruction technique for attenuation correction in high-resolu-tion positron emission tomography.Eur J Nucl Med,1996,23(6):656-661. 被引量:1
  • 7Wells RG,King MA,Simkin PH,et al.Comparing filteredbackprojection and ordered-subsets expectation maximization forsmall-lesion detection and localization in 67Ga SPECT.J NuclMed,2000,41(8):1391-1399. 被引量:1
  • 8Riddell C,Carson RE,Carrasquillo JA,et al.Noise reduction inoncology FDG PET images by iterative reconstruction:A quanti-tative assessment.J Nucl Med,2001,42(9):1316-1323. 被引量:1
  • 9张祺丰,彭芸,李剑颖,刘玥,曾津津.64排螺旋CT自动管电流调节技术降低儿童胸部扫描剂量的应用研究[J].中国医学影像技术,2009,25(9):1591-1594. 被引量:42
  • 10刘士远,于红.积极推进胸部低剂量CT扫描的临床应用[J].中华放射学杂志,2010,44(1):6-7. 被引量:81

共引文献74

同被引文献56

引证文献10

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部