超声造影在置管引流中的应用 被引量:1

Value of Contrast Enhanced Ultrasonography in Catheter Drainage
下载PDF
导出
摘要 目的探讨超声造影在置管引流后确定引流导管位置及引流导管是否通畅中的作用.方法回顾性分析超声引导下经皮穿刺置管引流后,经引流导管注射稀释后的超声造影剂Sonovue患者(78例,93支引流导管),观察显影后导管位置、引流效果.结果 81支引流导管全管清晰显影,导管头端位于预定位置,造影剂弥散区域与预定区域吻合,占87.1%;5支头端显影不清,但造影剂弥散区域与预定区域吻合,占5.4%;3支头端显影未能到达预定位置,4支管体部分显影,并突然截断,占7.5%,此7支引流导管拔出后再次置入并注射稀释后的超声造影剂,导管全管清晰显影,导管头端位于预定位置,造影剂弥散区域与预定区域吻合.结论经引流导管注射声学造影剂Sonovue造影对于引流导管显示、导管位置判断及引流区域判断方法可靠,无损伤,具有较高的临床应用价值. Objective To investigate the value of contrast enhanced ultrasonography in catheter drainage to determine the position and patency of catheter. Me thods We made a retrospective analysis in patients(78 cases)received percutaneous ultrasound-guided catheter drainage and diluted ultrasound contrast agents injection, and observed the catheter position and drainage effect. Re s ults Eighty-one drainage catheters developed the whole tube clearly, with catheter tips located at a predetermined position and the contrast agent diffusion region coincided with the predetermined region, accounting for 87.1%. Five catheter tips were unclear, while the contrast agent diffusion region coincided with the predetermined region, accounting for 5.4%. Three catheter tips failed to reach the predetermined position,and four drainage catheters developed part of the body of tubes clearly and showed cut off,accounting for 7.5%. Seven drainage catheter were pulled out and placed again, and then injected the diluted ultrasound contrast agent. In the replaced catheters, the catheters developed the whole tube clearly, catheter tips located at a predetermined position,and the contrast agent diffusion region coincided with the predetermined region.Conclus ion Contrast enhanced ultrasonography by diluted ultrasound contrast agent Sonovue injection through catheter can display drainage catheter,and determine the location of the catheter and drainage area. It is a reliable and non-invasive method and has a high clinical value.
作者 陈海涛 杨丽春 海赛苹 罗晓茂 CHEN Hai-tao,YANG Li-chun,HAI Sai-ping,LUO Xiao-mao(Dept. of Ultrsound,The Tumor Hospital of Yunnan Province,The Third Affiliated Hospital of Kunming Medical University,Kunming Yunnan 650118,China)
出处 《昆明医科大学学报》 CAS 2018年第4期62-65,共4页 Journal of Kunming Medical University
基金 国家自然科学基金资助项目(81560497)
关键词 超声造影 引流 稀释 超声造影剂 引流区域 通畅性 Contrast enhanced ultrasonography Drainage Diluted Ultrasound contrast agents Drainagearea Patency
  • 相关文献

参考文献17

二级参考文献160

共引文献137

同被引文献11

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部