期刊文献+

气道阻力及肺的顺应性对呼吸机压力控制精度的影响 被引量:7

Effect of Airway Resistance and Lung Compliance on the Pressure Control Precision of Ventilator
下载PDF
导出
摘要 呼吸机的稳定性、可靠性至关重要,但质控标准中并未对控制精度提出要求。本文设置了一种实验方法,验证了呼吸机在不同负载(不同的气道阻力和肺的顺应性)压力控制模式下精度情况。实验结果表明当呼吸力学参数改变时,呼吸机压力控制精度可能会受到影响,严重时可能超出企业标准,甚至影响呼吸机使用的安全性,造成患者肺损伤。 The stability and reliability of ventilator is of great signifcance. However, the standard of quality control does not claim any requirement for the control precision of ventilator. The present study designed an experiment to test the control precision of pressure control mode when the load of ventilator changed. The results showed that when the respiration parameters changed, the control precision of ventilator was in?uenced. It sometimes failed to meet corporate standard, even decreased the safety of ventilator, which might cause lung injury.
作者 侯晓旭 李佳戈 任海萍 HOU Xiaoxu;LI Jiage;REN Haiping(Institute for Medical Devices Control,National Institutes for Food and Drug Control,Beijing 102629,China)
出处 《中国医疗设备》 2018年第8期38-41,共4页 China Medical Devices
基金 国家重点研发计划(2016YFC0103201) 中国食品药品检定研究院中青年基金课题(2016C5)
关键词 气道阻力 肺的顺应性 压力控制模式(PCV) 呼吸机 airway resistance lung compliance pressure control ventilation (PCV) ventilator
  • 相关文献

参考文献9

二级参考文献36

共引文献27

同被引文献51

引证文献7

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部