期刊文献+

PL-GPC50凝胶色谱仪的构造及常见故障诊断处理 被引量:1

Common fault diagnosis and structure of PL-GPC50 gel chromatograph
下载PDF
导出
摘要 以PL-GPC50型凝胶色谱仪为例,介绍了凝胶色谱仪的结构、分析原理、常见故障现象及诊断、预防、处理方法。凝胶色谱仪一般分为流动相、泵系统、进样系统、色谱柱、检测系统、数据处理六大模块;凝胶色谱仪的故障通常发生在进样器、泵系统、色谱柱;流动相质量是凝胶色谱仪出现故障中的最大影响因素,需引起重视。 The structure, analysis principle, common fault phenomena , diagnosis, prevention and solution of gel chromato-graph were introduced by using PL-GPC50 gel chromatograph as an example .A gel chromatograph normally consists of six mod-ules, namely mobile phase, pump system, sampling system, chromatographic column, detection system, data processing.The faults of gas chromatogrph usually occur in the sampling system , pump system and chromatographic column .The mobile phase quality is the most important influential factor causing the faults , which should be paid more attention .
作者 苏凤仙 陈建梅 Su Fengxian Chen Jianmei(Research Institute ofSINOPEC Yizheng Chemical Fiber Co. , Ltd. , Yizheng 211900 Jiangsu Key Laboratory of High Performance Fiber, Yizheng 211900)
出处 《合成纤维工业》 CAS 2017年第5期75-78,共4页 China Synthetic Fiber Industry
关键词 凝胶色谱仪 结构 故障诊断 处理 预防 gel chromatograph structure fault diagnosis processing prevention
  • 相关文献

参考文献4

二级参考文献17

  • 1戴厚益,陈琪芳,杨华.高温凝胶渗透色谱法在聚苯硫醚树脂测试中的应用[J].塑料工业,2004,32(10):45-46. 被引量:1
  • 2白颖,李建伟.JASCO-1500凝胶色谱仪的维护与故障排除[J].分析测试技术与仪器,2006,12(1):59-61. 被引量:2
  • 3聂颖,崔小明.聚苯硫醚的生产及改性技术进展[J].化工科技市场,2006,29(9):9-14. 被引量:5
  • 4EhlersGFL,Fisch K R,Powell W R.Thermal degradation of polymers with phenylene units in the chain.Ⅱ.Sulfur-containing polyarylenes[J].J Polym Sci Part A Polym Chem,1969,7(10):2955-2967. 被引量:1
  • 5Reentz W D,Kaplan M L.Mass spectral analysis of poly(p-Phenylene sulphide)[J].Polymer,1982,23(2):310-313. 被引量:1
  • 6Brown C E,Kovacic P,Welch K J,et al.Laser desorption/foutier transform mass spectra of poly(phenylene sulfide),polyaniline,poly(vinyl phenol),polypyrene,and related oligomers:Evidence for carbon clusters and feasibility of physical dimension measurement[J].J Polym Sci Part A:Polym Chem,1988,26(1):131-148. 被引量:1
  • 7Peters O A,Still R H.The thermal degradation of poly(phenylene sulphide)-Part 1[J].Polym Degrad Stab,1993,42(1):41-48. 被引量:1
  • 8Peters O A,Still R H.The thermal degradation of poly(phenylene sulphide)-Part 2[J].Polym Degrad Stab,1993,42(1):49-53. 被引量:1
  • 9Budgell D B,Day M,Cooney J D.Thermal degradation of poly(phenylene sulfide)as monitored by pyrolysis-GC/MS[J].Polym Degrad Stab,1994,43(1):109-115. 被引量:1
  • 10Gies A P,Geibel J F,Hercules D M.MALDI-TOF MS study of poly(p-phenylene sulfide)[J].Macromolecules,2010,43(2):943-951. 被引量:1