摘要 以尼龙为基材、副品红氰化物为染料制备了辐射变色薄膜剂量计,并在60 Coγ射线及10 MeV电子束参考辐射场中对辐射变色薄膜的各剂量学响应性能进行了研究。研究结果表明,在1~210kGy剂量范围内,该辐射变色薄膜剂量计具有较好的剂量响应线性,且在本实验范围内未发现明显的能量和剂量率依赖性;该剂量计重复性好于1.0%,照射后置于低温条件下贮存2周内信号较稳定;在1~210kGy剂量线性范围内得到的辐射变色薄膜剂量计测量吸收剂量的扩展不确定度对60 Coγ射线为4.4%(k=2),对电子束为6.2%(k=2)。通过在加速器上进一步的实验表明,该辐射变色薄膜除可用于辐射加工中剂量监测外,还可用于电子束辐照参数的测量。 The radiochromic film dosimeter was prepared using nylon as substrate and pararosaniline cyanide as dye in this paper.In 60Coγ-ray and 10 MeV electron beam reference radiation fields,dosimetry response performance of the radiochromic film was studied.The results show that the dosimeter has the response curve of good linearity in the dose range of 1-210 kGy,and no distinct dependence on energy and dose rate in the experiment.The repeatability of the radiochromic film dosimeter is better than 1.0%.The signal of radiochromic film dosimeter after irradiation is stable under the condition of low temperature storage within 2weeks.In the dose range of 1-210 kGy,the extensive uncertainty of the radiochromic film dosimeter is 4.4%(k=2)for 60Coγ-ray irradiation and 6.2%(k=2)for electron beam irradiation.In addition,this dosimetry system was used to measure depth-dose distribution curve for 10 MeV electron beam,whichshows it's suitable and practicable for electron beam irradiation.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2016年第9期1680-1687,共8页 Atomic Energy Science and Technology
关键词 辐射加工 辐射变色薄膜 吸收剂量 Γ射线 电子束 radiation processing radiochromic film absorbed dose γ-ray electron beam
  • 相关文献

参考文献11

  • 1AKHAVAN A, SOHRABPOUR M, SHAR- IFZADEH M. Preparation of a new chemicalradiochromic film dosimeter[J]. Radiat Phys Chem, 2002, 63: 773-775. 被引量:1
  • 2KHAVAN A A, SOHRABPOUR M, SHAR IFZADEH M. Polyvinyl butyral films containing leuco-malachite green as low-dose dosimeters[J]. Radiat Phys Chem, 2008, 77: 457-462. 被引量:1
  • 3HUMPHERYS K C, KANTZ A D. Radiochro- mic: A radiation monitoring system[J]. Radiat Phys Chem, 1977, 9: 737-747. 被引量:1
  • 4MillerA,MclaughlinWL.对辐射变色染料薄膜和其他塑料剂量计在辐射加工条件下使用时的评价[R].维也纳:IAEA,1986:147. 被引量:2
  • 5对1978年全面比对结果的评述[R].维也纳:IAEA,1986. 被引量:1
  • 6MCLAUGHLIN W L, HUMPHREYS J C, LE- VINE H, et al. The gamma-ray response of radiochromic dye films at different absorbed dose rates[J]. Radiat Phys Chem, 1981, 18: 987- 992. 被引量:1
  • 7温祝堂,王永其,骆日佑,等.GF-型辐射变色尼龙薄膜剂量计的研究[J].计量学报,1988,9(4):304-308. 被引量:2
  • 8陈丽姝,闻友勤,孙存普.辐射变色薄膜电子高剂量计的一般特性[J].原子能科学技术,1989,23(6):1-8. 被引量:11
  • 9MCLAUGHLIN W L, KOSANIC M. The gain ma-ray response of pararosaniline cyanide dosi- meter solutions[J]. Appl Radiat Isot, 1974, 25: 249-262. 被引量:1
  • 10MCLAUGHLIN W L. Miroscopic visualization of dose distributions [J]. Appl Radiat Isot, 1966, 17: 85-96. 被引量:1

共引文献10

同被引文献67

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部