期刊文献+

柱温变化对高效液相色谱法测定黄芩苷的影响 被引量:5

Influence of temperature on parameters of HPLC by measuring Baicalin
下载PDF
导出
摘要 改变高效液相色谱(HPLC)分离柱的温度,对黄芩苷含量进行测定。随着温度的升高,黄芩苷的保留时间、峰面积、有效塔板数、分离度等参数都发生了相应变化变化,45℃时,系统压力为4.2 MPa,黄芩苷保留时间为5.193min,有效塔板数和分离度均最大,拖尾因子接近1.00,黄芩苷峰较窄且对称性好,图谱最优。 The content of baicalin was determined by HPLC while the separation column temperature was changed. The results showed that the parameters of HPLC, such as retention time, peak area, effective plate number and separating degree were correspondingly changed with the increase of temperature. At 45 ℃ ,the pressure of HPLC system was 4. 2 MPa,retention time was 5. 193, the trailing factor was close to 1.00, both the plate number and separating degree were all largest,the baicalin peak was narrow and symmetrical,and the chromatogram was optimal.
出处 《郑州牧业工程高等专科学校学报》 2015年第2期13-15,共3页 Journal of Zhengzhou College of Animal Husbandry Engineering
基金 河南省教育厅重点项目(13A61038) 河南省科技厅项目(132102110149)
关键词 高效液相色谱法 柱温变化 黄芩苷测定 影响 HPLC change in column temperature Baicalin measurement influence
  • 相关文献

参考文献8

  • 1中华人民共和国药典(一部)[S].北京:化学工业出版社,2010,4. 被引量:13
  • 2Y. Yang, A. D. Jones. Reerection behavior of phenols, anilines, and alkylbenzenes in liquid chromatographic separations using suberitical water as the mobile phase [J]. Anal Chem, 1999,71: 3808. 被引量:1
  • 3B. W. Yan, J. H. Zhao. High- temperature ultrafast liquid chromatography[J]. Anal Chem, 2000,72 : 1253-1262. 被引量:1
  • 4刘光会.色谱分析的新领域—高温液相色谱[J].天然气化工—C1化学与化工,2000,25(6):46-52. 被引量:12
  • 5J. W. Li, Y. Hu. Fast separations at elevated tempera- tures on polybutadiene- coated zirconia reversed- phase material[J]. Anal Chem, 1997 (69) :3884 - 3888. 被引量:1
  • 6W. S Hancock, R. C. Chloupek. Temperature as a varia- ble in reversed--phase HPLC separation of peptide and protein samples: selectivity effects observed in the sep- aration of several peptide and protein mixtures[J]. Chromatography A,1994(686):45-59. 被引量:1
  • 7T. Greibrokk, T. Andersen. High temperature liquid chromatography [J]. Chromatography A, 2003, 1000 (12) :743. 被引量:1
  • 8T. Kondo, Y. Yang. Comparison of elution strength, column efficiency, and peak symmetry in subcritical water chromatography and traditional reversed--phase liquid chromatography[J]. Analytical Chimica Acta, 2003(494) : 157-166. 被引量:1

二级参考文献3

  • 1G. Liu,Z. Xin. Retention study on silica gel in high-temperature liquid chromatography (HTLC)[J] 1998,Chromatographia(5-6):278~284 被引量:1
  • 2G. Liu,Z. Xin. The study of reversed phases by inverse gas chromatography[J] 1996,Chromatographia(5-6):290~294 被引量:1
  • 3G. Liu,N. Djordjevic,F. Erni. The application of ‘Z’-shaped UV cells in open-tubular capillary column liquid chromatography[J] 1994,Chromatographia(5-6):313~316 被引量:1

共引文献23

同被引文献36

引证文献5

二级引证文献9