期刊文献+

CdZnTe像素探测器的制备与表征 被引量:3

Fabrication and Characterization of CdZnTe Pixel Detector
下载PDF
导出
摘要 本文采用CdZnTe单晶制成像素探测器,并对其能谱响应特性及均匀性进行了系统表征。通过I-V和能谱响应测试,测定了晶体的电阻率和载流子迁移率与寿命的积,并用红外透过显微成像观察了晶体内Te夹杂的分布特性。采用光刻、剥离和真空蒸镀技术,在CdZnTe晶片上制备了8×8的像素电极,用丝网印刷和贴片技术通过导电银胶实现像素电极与读出电路的准确连接,制备出CdZnTe像素探测器。对像素探测器的测试表明,-300V下单像素最大漏电流小于0.7nA,对241 Am 59.5keV的能量分辨率可达5.6%,优于平面探测器。进一步分析了晶体内Te夹杂等缺陷对探测器漏电流和能谱响应特性的影响规律,结果表明,Te夹杂的聚集会显著增加漏电流,并降低探测器的能量分辨率。 The CdZnTe single crystal was fabricated into a pixel detector and its energy spectroscopic response and uniformity were systematically characterized.The resistivity and product of carrier mobility and lifetime were calculated by I-V and energy spectroscopic response test.The distribution of Te inclusions was observed with IR transmission imaging.A 8×8pixel electrode was fabricated employing photolithography,lift-off and electrode deposition techniques.Then,the detector was bonded to the readout PCB board with conductive silver adhesives using stencil printing and patching techniques.The testing results show that the maximum leakage current of single pixel is less than 0.7nA.The energy resolution for 241 Am 59.5keV is up to 5.6%,which is better than that of planar detector.The analysis of the influence of Te inclusions on leakage current and energy spectroscopic response predicts that the aggregation of Te inclusions will increase the leakage current and in consequence reduce the energy resolution of the detector.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2015年第7期1320-1324,共5页 Atomic Energy Science and Technology
基金 国家自然科学基金资助项目(61274081 51372205) 国家重大科学仪器设备开发专项资助(2011YQ040082)
关键词 CDZNTE 像素探测器 能量分辨率 CdZnTe pixel detector energy resolution
  • 相关文献

参考文献9

  • 1BUTLER J F, LINGREN C L, DOTY F P. Cdl-,. ZnxTe gamma ray detectors[J].IEEE Trans Nucl Sei, 1992, 39(4): 605-609. 被引量:1
  • 2刘志亮,毛用泽,邹士亚.CdZnTe探测器在天文物理研究中的应用综述[J].核电子学与探测技术,2009,29(1):1-5. 被引量:3
  • 3CARINI G A, BOLOTNIKOV A E, CAMAR- DA G S, et al. High-resolution X-ray mapping of CdZnTe detectors[J]. Nucl Instrum Methods Phys Res A, 2007, 579(1): 120-124. 被引量:1
  • 4BOLOTNIKOV A E, CAMARDA G S, CUI Y, et al. Characterization and evaluation of extended defects in CZT crystals for gamma-ray detectors [J]. J Cryst Growth, 2013, 379: 46-56. 被引量:1
  • 5BOLOTNIKOV A E, BOGGS S E, HUBERT- CHEN C M, et al. Properties of Pt Schottky type contacts on high-resistivity CdZnTe detec- tors[J]. Nucl Instrum Methods Phys Res A, 2002, 482(1): 395-407. 被引量:1
  • 6HECHT K. Zum mechanismus des lichtelektris- chen prim rstromes in isolierenden kristallen [J]. Zeitschrift far Physik, 1932, 77(3-4): 235-245. 被引量:1
  • 7AMMAN M, LEE J S, LUKE P N. Electron trapping nonuniformity in high-pressure-Bridg- man-grown CdZnTe[J]. J Appl Phys, 2002, 92 (6): 3 198-3 206. 被引量:1
  • 8BOLOTNIKOV A E, ABDUL-JABBER N, BABALOLA S, et al. Effects of Te inclusions on the performance of CdZnTe radiation detectors [C]//IEEE Nuclear Science Symposium Confer- ence Record. Dresden: IEEE, 2007: N27-2. 被引量:1
  • 9WILSON M D, CERNIK R, CHEN H, et al. Small pixel CZT detector for hard X-ray spectros- copy[J]. Nucl Instrum Methods Phys Res A, 2011, 652(1): 158-161. 被引量:1

二级参考文献18

  • 1http://www. evproducts. com/material_prop. pdf. 被引量:1
  • 2James L. Matteson, et al. CAZnTe arrays for astrophysics applications [ J ]. SPIE Proceedings, 1997,3115:160- 175. 被引量:1
  • 3Richard E. Rothschild, et al. Design and performance of the HEXIS CZT detector module[J]. Proceedings of SPIE, the International Society for Optical Engineering, 2002,9 : 45-53. 被引量:1
  • 4Goro Sato, et al. Properties of CdZnTe detectors in the Burst Alert Telescope (BAT) Array[J]. Proceedings of the SPIE, 2004, 5198 : 209-216. 被引量:1
  • 5T.O. T mer, et al. Progress in Detector Development for the All-sky X-ray & Gamma-ray Astronomy Monitor (AXGAM) [J]. Proceedings of the 26th International Cosmic Ray Conference, 1999, 5:112-115. 被引量:1
  • 6O. Limousin. New trends in CdTe and CAZnTe detectors for X- and gamma-ray applications[J]. Nuclear Instruments and Methods in Physics Research Section A, 2003,504(1-3) : 24-37. 被引量:1
  • 7J. Tueller, et al. InFOCuS Hard X-ray Imaging Telescope[J]. Experimental Astronomy, 2005,20 (1-3) : 121-129. 被引量:1
  • 8C. M. Hubert Chen, et al. characterization of the HEFT CdZnTe pixel detectors[J]. Proceedings of the SHE, 2004,5198:9-18. 被引量:1
  • 9J. N. Goswami,et al. High energy X-γ ray spectrometer on the Chandrayaan-1 mission to the Moon [J]. Journal of Earth System Science, 2005,114(6) : 733-738. 被引量:1
  • 10C.M. Stahleb, et al. CdZnTe detectors for gamma-ray Burst ArcSecond Imaging and Spectroscopy (BASIS)[J]. Nuclear Instruments and Methods in Physics Research A, 1996, 380:486-489. 被引量:1

共引文献2

同被引文献22

  • 1黄勇,杨磊,杨汝基,大野良一.碲化镉(CdTe)探测器的原理及医学应用[J].上海生物医学工程,2005,26(4):221-225. 被引量:12
  • 2闵嘉华,桑文斌,刘洪涛,钱永彪,滕建勇,樊建荣,李万万,张斌,金玮.热处理方法改善CdZnTe晶体性能研究(英文)[J].稀有金属材料与工程,2007,36(3):471-474. 被引量:3
  • 3XU Huichao,CHENG Cheng,ZHAO Cuilan,ZHANG Jinzhou,PU Shijie.Characteristics of a prototype CdZnTe detector[J].Nuclear Science and Techniques,2007,18(6):358-361. 被引量:3
  • 4CHAUDHURI S K, KRISHNA R M, ZAVALLA K J, et al. Cd0.gZn0. Te crystal growth and fabrication oI large volume single polarity charge sensing gamma detectors [J]. IEEE Trans Nucl Sci, 2013, 60(4) : 2 853 - 2 858. 被引量:1
  • 5WEI G L, WANG L J, ZHANG J J, et al. Growth and char- acterization of indium-doped Cd ZnTe crystal by traveling heater method[J]. Nucl Instrum Methods A, 2013, 704:127 - 130. 被引量:1
  • 6NIU L B, LI Y L, ZHANG L, et al. Performance simulation and structure design of hinode CdZnTe gamma-ray detector [J]. Nuclear Science and Techniques, 2014, 25(1) : 010406. 被引量:1
  • 7CAROLI E, CURADO DA SILVA R M, STEPHEN J B, et aI. A polarimetric experiment with a laue lens and CZT pixel detector[J]. IEEE Trans Nucl Sci, 2009, 56(4): 1 848- 1 854. 被引量:1
  • 8GUILLAUME M, LUX S, MONNET O, et al. Studying spatial resolution of CZT detectors using sub-pixel positioning for SPECT[J]. IEEE Trans Nucl Sci, 2014, 61(5): 2 559 -2 566. 被引量:1
  • 9KRISI-tNA R M, CHAUDHURI S K, ZAVALLA K C, et al. Characterization of Cdo.9 Zn0.Te based virtual Friseh grid detectors for high energy gamma-ray detection [J]. Nucl In strum Methods A, 2013, 701: 208-213. 被引量:1
  • 10CHAUDHURI S K, NGVYEN K, PAK R O, et al. Large area Cd0.gZno. Te pixelated detector: fabrication and charac terization[J]. IEEE Trans Nucl Sci, 2014, 61(2) : 793 -798. 被引量:1

引证文献3

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部