应用两种三维探测器阵列进行螺旋断层调强放疗计划剂量验证 被引量:2

Dose Verification of Helical Tomotherapy in the Intensity Modulated Planning with Three-dimensional Semiconductor Arrays Delta^4 and ArcCheck
下载PDF
导出
摘要 目的:研究分别采用Delta4和Arc Check三维探测器阵列对螺旋断层放疗系统(TOMO)的调强计划实施剂量验证的异同。方法:分别采用Scandi Dos公司Delta4三维半导体探测器阵列和Sun Nuclear公司Arc Check三维半导体探测器阵列对10例患者的TOMO调强放疗计划实施验证。通过TOMO系统MVCT成像和配准消除床面沉降的影响,确保实现两种三维探测器阵列的精确摆位。实施治疗质量保证(Delivery Quality Assurance,DQA)计划照射后将探测器阵列测量获得的平面剂量分布与TOMO计划系统模体计划计算获得的平面剂量分布进行比较。采用Gamma分析方法,选择多个不同的剂量标准分析评估验证情况。结果:阈值水平(Threshold,TH)分别为5%、10%、20%、70%、90%时,采用3%/2 mm、3%/3 mm、3%/4 mm、3%/5 mm四种不同剂量偏差/吻合距离剂量标准,γ≤1的平均通过率均超过94.1%,测量所得剂量分布与计算结果在相应平面的几何分布均呈现出良好的一致性。结论:两种三维探测器阵列各具特点,用于TOMO调强放疗计划验证能够获得令人满意的结果 。 Objective To compare the Delta4 three dimensional detector array with the ArcCheck detector array in dose verifica- tion of tomotherapy (TOMO) intensity modulated planning. Methods 10 patients were selected to verify dose distribution in the TOMO plan with the Delta4 and ArcCheck detector arrays. The precise setup of detector arrays was insured by MVCT imaging. The measured dose distribution on the corresponding planes were compared with those calculated by the treatment planning sys- tem (TPS) after the implementation of delivery quality assurance (DQA) plan. Several different dose distribution assessments were chosen to verify the results made by Gamma analysis. Results 4 different doses 3%/2 mm, 3%/3 mm, 3%/4 mm, 3%/5 mm were used when the thresholds were THs, THe0, THe, THTo, THg0. The mean passing rates with γ parameter ≤ 1 were all above 94.1%. The dose distribution measured by these two kinds of three dimensional detector arrays combining corresponding phan toms was well consistent with which was calculated by TPS. Conclusion Three dimensional detector arrays both the Delta4 and ArcCheck with corresponding phantoms can get satisfying results when used in dose verification of TOMO plan.
出处 《中国医学物理学杂志》 CSCD 2015年第2期218-220,238,共4页 Chinese Journal of Medical Physics
基金 全军医学计量专项课题(2011-JL2-005)
关键词 螺旋断层放疗 三维探测器阵列 调强放疗 剂量验证 质量保证 tomotherapy three dimensional detector array IMRT dose verification quality assurance
  • 相关文献

参考文献21

  • 1Welsh JS, Lock M, Harari PM, et al. Clinical implementation of adaptive helical tomotherapy: a unique approach to image-guided intensity modulated radiotherapy[J]. Technol Cancer Res Treat, 2006, 5: 465-79. 被引量:1
  • 2Duma MN, Kampfer S, Schuster T, et al. Do we need daily image-guided radiotherapy by megavoltage computed tomography in head and neck helical tomotherapy? The actual delivered dose to the spinal cord[J]. Int J Radiat Oncol Biol Phys, 2012, 84: 283-8. 被引量:1
  • 3Voordeckers M, Farrag A, Everaert H, et al. Parotid gland sparing with helical tomotherapy in head-and-neck cancer[J]. Int J Radiat Oncol Biol Plays, 2012, 84: 443-8. 被引量:1
  • 4Langen KM, Papanikolaou, Balog J, et al. QA for helical tomotherapy: Report of the AAPM task group 148 [J]. Med Phys, 2010, 37: 4817-4853. 被引量:1
  • 5Sarninathan S, Maniekan R, Chandraraj V, et al. Dosimetric study of 2D ion chamber array matrix for the modem radiotherapy treatment verification[J]. J Appl Clin Med Phys, 2010, 11: 116-127. 被引量:1
  • 6Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine [J]. Phys Med Biol, 2007, 52:1197-1208. 被引量:1
  • 7Woo M, Nico A. Using a matrix detector for rotational delivery QA[J]. Med Phys, 2007, 34: 2502. 被引量:1
  • 8Chandraraj V, Stathakis S, Manickam R, et al. Comparison of four commercial devices for RapidArc and sliding window IMRT QA[J]. J Appl Clin Med Plays, 2011, 12: 338-349. 被引量:1
  • 9Feygelman V, Opp D, Javedan K, et al. Evaluation ofa 3D diode array dosimeter for helical tomotherapy delivery QA [J]. Med Dosimetry, 2010, 35: 324-329. 被引量:1
  • 10Van Esch A, Clermont C, Devillers M, et al. On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the octavius phantom [J]. Med Phys, 2007, 34: 33825-3837. 被引量:1

二级参考文献36

  • 1Woo M, Nico A. Using a matrix detector for rotational delivery QA. Med Phys ,2007 ,34 :2502. 被引量:1
  • 2Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel-segmented ionization chamber. Med Phys ,2004,31:414-420. 被引量:1
  • 3Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys ,2003,30 : 2455-2464. 被引量:1
  • 4Nelms BE, Simon JA. A survey on planar IMRT QA analysis. J Appl Clin Med Phys,2007 ,8 :876-890. 被引量:1
  • 5Koch N,Ashenafi M,Li J,et al. Evaluation of a 2D ion chamber array for Quality Assurance of Intensity Modulated Arc Therapy. Med Phys,2008,35 : 2945. 被引量:1
  • 6Araki F, Tajiri S, Tominaga H, et al. Application of a 2-D ion chamber array for dose verification of dynamic IMRT with a micro- MLC. Int J Radiat Oncol Biol Phys ,2006,66:703-704. 被引量:1
  • 7Stasi M, Giordanengo S, Cirio R, et al. D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer. Phys Med Biol,2005 ,50 :4681-4694. 被引量:1
  • 8Mackie TR,Holmes T,Swerdloff S,et al.Tomotherapy:a new concept for the delivery of conformal radiotherapy[J].Med Phys,1993,20:1709-1719. 被引量:1
  • 9Mackie TR,Balog J,Ruchala K,et al.Tomotherapy[J].Semin Radiat Oncol,1999,9:108-117. 被引量:1
  • 10Yang JN,Mackie TR,Reckwerdt P,et al.An investigation of tomotherapy beam delivery[J].Med Phys,1997,24:425-436. 被引量:1

共引文献13

同被引文献22

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部