TomoTherapy治疗计划剂量验证方法的研究 被引量:2

Research on Dose Verification Method for Tomo Therapy Plan
下载PDF
导出
摘要 目的:建立一种可以对TomoTherapy治疗计划剂量进行验证的方法。方法:根据IAEA TRS 277号报告提供的水下吸收剂量测量计算的基本原理,采用待检的TomoTherapy辐射源以及在权威计量部门检定过的0.65 cc Farmer电离室,参照交叉校准的方法将标准量值传递给0.015 cc针尖电离室测量系统。之后采用固体水球型模体对治疗计划剂量进行检测和验证。结果:采用所建立的方法,剂量检测结果与治疗计划剂量相对偏差分别为1.35%、-2.30%、2.88%和0,皆小于±3%。结论:采用本研究所建立的方法可以与设备生产厂家的剂量治疗计划、检测结果进行比对、核查和控制,对于TomoTherapy项目的防护与应用质量检测具有重要的现实意义。 Objective To establish a method used for verifying TomoTherapy plan dose.Methods According to the principle about measuring the underwater point absorbed dose provided in the IAEA Report No.277,0.65 cc Farmer ion chamber corrected by authoritative department was used to measure the absorbed dose.Then the measurement result was transferred to 0.015 cc pinpoint by means of cross correction.A spheroid solid water phantom was set up as a DQA phantom to verify the accuracy of TomoTherapy plan point dose.Result The relative difference between the measured and calculated point doses was 1.35%,-2.30%,2.88%,0,respectively,conforming to the criteria of ±3%.Conclusion The method established in the study could be used for examining and comparison with the measurement protocol provided by the manufacturers.
出处 《医疗卫生装备》 CAS 2013年第7期105-107,共3页 Chinese Medical Equipment Journal
基金 总后卫生部军事医学计量立项课题(2011-JL2-005)
关键词 TOMOTHERAPY 剂量 治疗计划系统 测量 TomoTherapy dose plan dose measurement
  • 相关文献

参考文献11

二级参考文献37

  • 1高磊,罗素明,邱杰,何志坚,杨波,庞庭田.金属氧化物半导体场效应管探测器在放射治疗中的剂量学研究[J].中华放射医学与防护杂志,2008,28(2). 被引量:1
  • 2耿辉,戴建荣,李晔雄,张可,王绿化.一种简单调强放疗技术应用的初步研究[J].中华放射肿瘤学杂志,2006,15(5):411-415. 被引量:58
  • 3Woo M, Nico A. Using a matrix detector for rotational delivery QA. Med Phys ,2007 ,34 :2502. 被引量:1
  • 4Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel-segmented ionization chamber. Med Phys ,2004,31:414-420. 被引量:1
  • 5Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys ,2003,30 : 2455-2464. 被引量:1
  • 6Nelms BE, Simon JA. A survey on planar IMRT QA analysis. J Appl Clin Med Phys,2007 ,8 :876-890. 被引量:1
  • 7Koch N,Ashenafi M,Li J,et al. Evaluation of a 2D ion chamber array for Quality Assurance of Intensity Modulated Arc Therapy. Med Phys,2008,35 : 2945. 被引量:1
  • 8Araki F, Tajiri S, Tominaga H, et al. Application of a 2-D ion chamber array for dose verification of dynamic IMRT with a micro- MLC. Int J Radiat Oncol Biol Phys ,2006,66:703-704. 被引量:1
  • 9Stasi M, Giordanengo S, Cirio R, et al. D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer. Phys Med Biol,2005 ,50 :4681-4694. 被引量:1
  • 10IAEA. Technical Reports Series No. 277. Absorbed dose determination in photon and electron beams. Vienna: IAEA, 1997. 被引量:1

共引文献26

同被引文献23

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部