Sirius和Lenstar及A超测量中央角膜厚度的对比研究 被引量:1

Comparative study of central corneal thickness measurements with Sirius,Lenstar and Ultrasound pachymetry
下载PDF
导出
摘要 目的比较Sirius、Lenstar及A超测量中央角膜厚度(CCT)的差异,并对测量结果进行一致性评价。方法对52例(104眼)近视眼患者,依次采用Sirius、Lenstar及A超测量CCT。配对t检验和Pearson相关分析评价CCT测量值两两间的差异及相关关系,Bland-Altman分析评价一致性。结果 Sirius与A超、Lenstar与A超的CCT测量值之间差异均有统计学意义,而Sirius与Lenstar的CCT测量值比较差异无统计学意义。三种仪器测量CCT具有显著的正相关性。CCT测量值95%一致性界限分别为Sirius与A超(-16.7,8.0)μm、Lenstar与A超(-19.7,8.5)μm、Sirius与Lenstar(-12.8,15.3)μm。结论 Sirius、Lenstar与A超测量CCT之间具有较高的一致性,可作为角膜中央厚度测量的常规工具。Sirius和Lenstar测量值与A超相比,存在差异,前二者低估了中央角膜厚度。 Objective To assess the differences and agreements among measurements of central corneal thickness (CCT) using Sirius, Lenstar and ultrasound pachymetry (USP) . Methods The CCT of 104 eyes from 52 patients with myopia were measured with Sirius, Lenstar and ultrasound pachymetry. Data were compared by using paired t-tests and the Pearson correlations. Bland-Altman analysis of all pairs were determined. Results The difference between Sirius and USP, Lenstar and USP were statistically significant (P 〈 0. 001 ), but the CCT measured by Sirius and Lenstar had no statistically significant difference (P 〉 0.05 ) . There was high correlation between the three instruments. The 95% limits of agreement for central cor- neal thickness measurements were - 16. 7μm to + 8. 0μm between Sirius and USP, - 19.7μm to + 8. 5μm between Lenstar and USP, and - 12. 8μm to + 15.3μm between Sirius and Lenstar. Conclusion Central corneal thickness obtained with the USP were similar to those obtained with both the Sirius and Lenstar. They all can be used as a routine tool of central corneal thickness measurement. But compared with USP, Sirius and Lenstar may underestimate the central corneal thickness.
出处 《广州医药》 2014年第6期14-17,共4页 Guangzhou Medical Journal
基金 广州市医药卫生科技项目(20131A011027)
关键词 A超 中央角膜厚度 Sirius Lenstar Ultrasound pachymetry Central corneal thickness
  • 相关文献

参考文献8

二级参考文献50

  • 1高磊,王忠浩,范慧君,谭丽霞.眼用表面麻醉液对近视眼角膜厚度的影响(英文)[J].国际眼科杂志,2005,5(3):428-432. 被引量:6
  • 2朱乐如,查屹,施明光.两种角膜地形图系统测量正视儿童角膜的比较[J].眼视光学杂志,2007,9(2):120-123. 被引量:4
  • 3陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:217
  • 4Terry M A, Ousley P J. Replacing the endothelium without corneal sur- face incisions or sutures : the first United States clinical series using the deep lamellar endothelial keratoplasty procedure [ J ]. Ophthalmology, 2003,110(4) : 755-764. 被引量:1
  • 5Price F J, Price M O. Descemet's stripping with endothelial keratoplasty in 50 eyes: a refractive neutral corneal transplant [J]. J Refract Surg, 2005,21 (4) :339-345. 被引量:1
  • 6Harris J E, Nordquist L T. The hydration of the cornea. I. The transport of water from the cornea [ J 1. Am J Ophthalmol, 1955,40 (5 Part 2) : 100-110. 被引量:1
  • 7Marsich M W, Bullimore M A. The repeatability of corneal thickness measures[ J ]. Cornea, 2000, 19(6) : 792-795. 被引量:1
  • 8Chang D T,Pantcheva M B,Noecker R J. Corneal thickness and in- traocular pressure in edematous corneas before and after Descemet stripping with automated endothelial keratoplasty [ J ]. Cornea, 2010,29 (10) : 1125-1130. 被引量:1
  • 9Whitacre M M, Stein R A, Hassanein K. The effect of corneal thickness on applanation tonometry [J]. Am J Ophthalmal,1993,115 (5): 592-596. 被引量:1
  • 10Emara B, Probst L E, Tingey D P, et al. Correlation of intraocular pres- sure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis[ J ]. J Cataract Refract Surg, 1998,24(10) : 1320-1325. 被引量:1

共引文献13

同被引文献12

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部