摘要
介绍了一个旨在将高时间分辨率和高探测灵敏度相结合、用于惯性约束聚变 (ICF)研究的时间分辨的中子诊断技术的中子示波管 ,初步实验获得了技术时间分辨约 40 ps,可探测中子总产额 10 8。
This article presents an oscillograph with both high time resolution and detection efficiency used in time-resolved neutron diagnosis in ICF (inertial confinement fusion). The oscillograph, comprising a neutron cathode, a specially designed electron optical configuration and a pair of plane deflector, is similar to a streak camera in principle except that no spatial information can be obtained. A sandwich-like polyethylene-CsI-aluminium is used to convert neutron into electron beam. The electron optics system converges the electron beam excited by neutron from a wide area at the image plane. These electrons are steered transversely by a linear voltage to obtain the intensity evolution versus time. In the experiment, a neutron generator was used to test the function and the detective efficiency of the neutron cathode, resulting in a neutron beam of approximately 10 8 be recorded in total flux. Also an equivalence of the oscillograph with a multi-alkali photocathode was used to simulate its operation in dynamic mode by which the instrumental resolution is calibrated to be 40ps around in the proof-of-principle experiment.
出处
《强激光与粒子束》
EI
CAS
CSCD
北大核心
2002年第4期633-636,共4页
High Power Laser and Particle Beams