摘要 目的:评价磁共振相位对比法(PC MR)测量主动脉血流的准确性。方法:由2位医师先后2次对11例正常志愿者的左心室及主动脉起始部的MR数据进行测量、分析,获得主动脉起始部心动周期内的净血流量(即每搏输出量,SV),并与左心室短轴位MR电影中左心室舒张期和收缩期容积的改变(SV)进行比较,探讨PC MR法测量主动脉血流的准确性;并观察上述两种方法在检查及后处理过程中所花费的时间,以及受测量者个体的影响情况。结果:PC MR法测量主动脉起始部所获得左心室的SV与左心室短轴位MR电影测量的结果具有良好的一致性(r=0.92);PC MR法测量SV花费时间较少、受测量者的影响小。结论:PC MR法能准确测量主动脉的血流,并能快速获得SV等左心室功能评价的指标。 Objective:To evaluate the accuracy of phase contrast (PC) MR in measuring the blood flow of aorta.Method:Eleven cases of blood flow in the beginning of aorta of the healthy volunteers were measured by using PC MR,and the stroke volume which was derived from the change of left ventricle within each cardiac cycle were measured by short axis cine MRI.The values of stroke volume by the two methods were analysed and compared.We compared the time that cost in scanning and post-processing of the two methods.Results:There were good consistency between net blood flow volume of aorta measured by using PC MR with that of left ventricle by short axis cine MRI (r=0.92).It took less time to scan and post-process imaging by using PC MR than short axis cine MRI.There was less inter-observer and intra-observer influence by using PC MR.Conclusion:PC MR is a noninvasive,safe,accurate method.It is a good way in research the blood flow dynamics of aorta by using PC MR,and can assess left ventricular systolic function.
出处 《中国临床医学影像杂志》 CAS 2014年第7期482-485,共4页 Journal of China Clinic Medical Imaging
基金 教育部博士点基金(新教师类 20115317120003) 云南省教育厅重点项目(2012Z091)
关键词 主动脉 磁共振成像 Aorta Magnetic resonance imaging
  • 相关文献

参考文献11

  • 1张镜如.生理学[M].4版.北京:人民卫生出版社,2000:231. 被引量:3
  • 2徐承煮,曹承刚.人体结构功能学[M] .北京:中国协和医科大学出版社,2003:346-374. 被引量:1
  • 3Meier D,Maier S,Bosiger P.Quantitative flow measurements on phantoms and on blood vessels with MR[J] .Magn Reson Med,1988,8(1):25-34. 被引量:1
  • 4O'Brien KR,Myerson SG,Cowan BR,et al.Phase contrast ultrashort TE:a more reliable technique for measurement of highvelocity turbulent stenotic jets[J] .Magn Reson Med,2009,62(3):626-636. 被引量:1
  • 5刘明,李彩英,周立霞,张铁亮,刘辉,刘怀军,全冠民.二维相位对比MRI在稳态流体模型的定量测量[J].放射学实践,2009,24(3):247-250. 被引量:7
  • 6Kondo C,Caputo GR,Semelka R,et al.Right and left ventricular stroke volume measurements with velocity-encoded cine MR imaging:in vitro and in vivo validation[J] .AJR,1991,157(1):9-16. 被引量:1
  • 7James SH,Wald R,Wintersperger B J,et al.Accuracy of right and left ventricular functional assessment by short-axis vs axial cine steady-state free-precession magnetic resonance imaging:intrapatient correlation with main pulmonary artery and ascending aorta phase-contrast flow measurements[J] .Can Assoc Radiol J,2013,64(3):213-219. 被引量:1
  • 8闫钟钰,李坤成,李永忠,周怀琪,杜富会.MR相位对比法和形态体积分析法评价心室功能的比较研究[J].中华放射学杂志,2005,39(3):262-266. 被引量:7
  • 9Nayler DL,Firmin DN,Longmore DB.Blood flow imaging by cine magnetic resonance[J] .J Comput Assist Tomogr,1986,10(5):715-722. 被引量:1
  • 10Sakuma H,Fujita N,Foo TK,et al.Evaluation of left ventricular volume and mass with breath hold cine MR imaging[J] .Radiology,1993,188(2):377-380. 被引量:1

二级参考文献28

  • 1李彩英,田建明,刘辉,陆建平,黄勃源,刘怀军.利用磁共振相位对比法对非搏动流体模型的定量测量研究[J].中国医学影像技术,2006,22(4):535-537. 被引量:10
  • 2李彩英,黄勃源,崔彩霞,刘怀军,孟令惠.MRI电影相位对比与超声对照对正常志愿者餐前、餐后门静脉血流定量研究[J].放射学实践,2007,22(5):471-474. 被引量:6
  • 3Iotz J, Meier C, Leppert A, et al. Cardiovascular Flow Measure ment with Phase-contrast MR Imaging; Basic Facts and Implementation[J]. RadioGraphics,2002,22(3) :651-671. 被引量:1
  • 4Greil G,Geva T, Maier SE, et al. Effect of Acquisition Parameters on the Accuracy of Velocity Encoded Cine Magnetic Resonance Imaging Blood Flow Measurements[J].Magn Reson lmaging, 2002,15(1) :47-54. 被引量:1
  • 5Bakker CJ, Hoogeveen RM, Viergever MA. Construction of a Protocol for Measuring Blood Flow by Two-dimensional Phase -contrast MRA[J]. J Magn Reson Imaging,1999,9(1):119-127. 被引量:1
  • 6Buonocore MH, Bogren H. Factors Influencing the Accuracy and Precision of Velocity-encoded Phase Imaging [J].Magn Reson Med,1992,26(1):141-154. 被引量:1
  • 7Tang C,Blatter DD, Parker DL. Accuracy of Phase-contrast Flow Measurements in the Presence of Partial-volume Effects[J]. J Magn Reson Imaging, 1993,3 (2) : 377-385. 被引量:1
  • 8Stahlbe W, Sondergaard L, Thomsen C, et al. Quantification of Complex Flow Using MR Phase Imaging: a Study of Parameters Influency the Phase velocity Relation[J]. Magn Reson Imaging, 1992,10(1) :13-23. 被引量:1
  • 9Burkart DJ,Johnson CD, Ehman RL, et al. Evaluation of Portal Venous Hypertension with Cine Phase-contrast MR Flow Measurements: High Association of Hyperdynamic Portal Flow with Variceal Hemorrhage[J]. Radiology, 1993,188(3):643-648. 被引量:1
  • 10Lorie R,Norbert J, Stephen C, et al. Arterial and Venous Blood Flow: Noninvasive Quantitation with MR Imaging[J]. Radiology, 1992,185(3):809-812. 被引量:1