摘要 本文研制开发了测量复合树脂固化收缩量的实验装置和检测方法。该装置采用激光测距传感器(测量精度为 1μm)和特制的材料固化腔 ,可以直接在光固化机照射的同时 ,测量复合树脂固化收缩的全过程。研究中采用 3种光固化机对 4种复合树脂材料进行 2种照射时间的固化收缩实验。经对实验结果分析发现 :复合树脂的固化收缩量的大小与固化光源的种类有关 ,用高能氙灯光源照射产生的固化收缩量最小 ;采用高能氙灯光源照射时固化收缩量小 ,并不完全是由于照射时间短 ,使复合树脂固化反应不充分所致 ,而是与光源本身的特性有关。 Objectives: The purpose of this study was to compare the effects of kind of light source on the polymerization shrinkage mass of composite resin. Methods: A measure equipment was developed, a displacement sensor was used (the accuracy is 1μm). The shrinkage mass of composite resin in a polymerization box can be direct measure, and the all process of shrinkage can be measured from the beginning of irradiation of light source. Three kinds of light source and four kinds of composite resin were tested with two irradiation times. Results: The polymerization shrinkage masses relate to the kinds of light source and the shrinkage mass produced by the high power xenon light source is least. Conclusion: The causation of this result is not always by dueto shorter time of irradiation and by the polymerization reaction insufficiency, but related to the character of the curing light source.
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 2002年第3期246-250,共5页 Chinese Journal of Biomedical Engineering
关键词 复合树脂 固化收缩量 固化光源 牙科材料 Composite resin Polymerization shrinkage mass Curing light source
  • 相关文献

参考文献8

  • 1[1]Phillips RW. Science of dental materials[M]. Washington; W.B.Saunders Tents. EDRTION company.1996,292-294. 被引量:1
  • 2[2]Tanoue N, Matsumura H, et al. Properties of four composite veneering materials polymerized with different laboratory photo-curing units[J]. Oral Rehabilitation. 1998,25(5):358-364. 被引量:1
  • 3[3]Sfondrini MF, Cacciafesta V, et al. Effects of conventional and high-intensity light-curing on enamel shear bond strength of composite resin and resin-modified glass-ionomer[J]. Amer J Orthodon Dentof Orthoped. 2001,119(1):30-35. 被引量:1
  • 4[4]国际标准ISO 4049-2000 DentistryPolymer-based filling restorative and luting materials[S]. 被引量:1
  • 5[6]Watts DC, Hindi AAL. Intrinsic`soft-start' polymerisation shrinkage-kinetics in an acrylate-based resin-composite[J]. Dent Materials,1999,15(1):39-45. 被引量:1
  • 6平泽忠.治疗用照射器の特性[J].The journal of dent engineer,1999,(130):1-5. 被引量:1
  • 7稻井纪通等.高出力照射器とレジンの硬化[J].J dental engineer,2000,(132):13-18. 被引量:1
  • 8[9]de Dee AJ, Feilzer AJ, et al. True linear polymerization shrinkage of unfilled resins and composites determined with a linometer[J]. Dent Materials.1993,9(1):11-14. 被引量:1