利用常规模拟机测量和离线校正体部肿瘤的摆位误差

Measurement and Off-line Modification of Set-up Errors in Body Tumors with Simulator
原文传递
导出
摘要 [目的]应用常规模拟机测量和离线修正体部肿瘤患者的首次摆位误差。[方法]应用瓦里安模拟治疗机对138例体部肿瘤患者在治疗前拍摄正侧位影像,通过与治疗计划的DRR图像匹配,测出体部肿瘤患者首次的摆位误差并进行修正。[结果]138例患者首次摆位有81次误差大于3mm,需要移动治疗床调整。胸、腹部肿瘤校正比例分别为45/71和36/67。胸部肿瘤摆位误差X轴(-0.09±2.98)mm、Y轴(-0.38±4.63)mm、Z轴(0.95±3.97)mm,腹部肿瘤摆位误差X轴(0.40±1.90)mm、Y轴(-0.70±3.20)mm、Z轴(-0.80±2.20)mm。校正后摆位误差缩小。各方向误差值中,Y轴误差值最大。[结论]利用常规模拟机测量并修正首次摆位误差是一种简便有效的验证方法。 [ Purpose ] To measure and modify the first time off-line set-up errors in body tumors with simulator. [ Methods ] One hundred and thirty-eight patients with body tumors were treated by Varian Acuity simulator. All patients were immobilized with fixation devices. DRR of each patient compared with portal images from simulator and set-up errors were recorded and ana- lyzed. [Results] One hundred and thirty-eight cases of first set up had 81 times error greater than 3mm,which need to move and adjust the treatment bed. The correction ratio of chest tu- mor and abdominal tumor were 45/71 and 36/67 respectively. The systemic+random errors in chest tumor on X,Y,Z axes were (-0.09±2.98)mm, (-0.38±4.63)mm and (0.95±3.97)mm,respec- tively,and in abdominal tumor on X,Y,Z axes were (0.40±1.90)mm, (-0.70±3.20)mm,(-0.80± 2.20)mm respectively. Set-up errors reduced after modification. Error of Y axis was the largest one. [ Conclusions ] The simulator for position calibration can measure and modify the first time set-up error. It is effective and convenient location verification.
机构地区 慈溪市人民医院
出处 《肿瘤学杂志》 CAS 2014年第4期323-325,共3页 Journal of Chinese Oncology
关键词 常规模拟机 体部肿瘤 首次摆位误差 离线修正 simulator body tumors first time set-up error off-line modification
  • 相关文献

参考文献6

二级参考文献38

  • 1王鑫,胡超苏,应红梅.放射治疗中的摆位误差[J].肿瘤防治杂志,2004,11(11):1221-1225. 被引量:32
  • 2段诗苗,江浩.鼻咽癌放疗面罩固定技术方法的探讨[J].蚌埠医学院学报,2007,32(2):154-155. 被引量:7
  • 3Baroni G,Ferrigno G,Orecchia R,et al.Real time three dimentional motion analysis for patient positioning verification[J].Radiother Oncol,2000,54(1):21-27. 被引量:1
  • 4Samuelsson A,Mercke C,Johansson K A. Systematic set-up er rors for IMRT in the head and neck region: effect on dose distribution[J]. Radiother Oncol,2003, 66(3) :303-311. 被引量:1
  • 5Beckham W A, Keall P J, Siebers J V. A fluence convolution method to calculate radiation therapy dose distributions that incorporate random set-up error[J]. Phys Med Biol, 2002,47 (19):3465-3473. 被引量:1
  • 6Stroom J C,de Boer J C,Huizenga H,et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probility[J]. Int J Radiat Oncol Biol Phys, 1999,43(4):905-919. 被引量:1
  • 7Van Herk M,Remeijer P,Rasch C,et al. The probility of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy[J]. Int J Radiat Oncol Biol Phys,2000,47(4) :1121-1135. 被引量:1
  • 8van Lin E N, van der Vight L, Huizenga, et al. Set-up improvement in head and neck radiotherapy using a 3D off line EPID based correction protocol and a customised head and neck sup port[J]. Radiother and Oncol,2003,68(2): 137-148. 被引量:1
  • 9Mock U, Dieckmann K, Wolff U, et al. Portal imaging based definition of the planning target volume during pelvic irradiation for gynecological malignancies[J]. Int J Radiat Oncol Biol Phys,1999,45(1) :227-232. 被引量:1
  • 10Tsai J S, Engler M J, Ling M N, et al. A non-invasive immobiliza tion. system and related quality assurance for dynamic intensity modulated. radiation therapy of intracranial and head and neck disease[J].Int J Radiat Oncol Biol Phys, 1999,43 (2) :455 - 467. 被引量:1

共引文献37

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部