应用模拟机对头颈肿瘤适形调强放疗进行位置校准探讨 被引量:3

Application of Simulator for position calibration in conformal and IMRT of head and neck cancer
下载PDF
导出
摘要 目的:应用模拟定位机对头颈肿瘤适形调强放疗进行位置校准。方法:应用模拟机对36例头颈肿瘤适形调强放疗进行位置校准后拍摄校准片,放疗前在加速器上同一体位再摄验证片1次。选择典型骨性标志作为对比参考点,模拟机校准片和加速器验证片分别与来自Pinnacle3治疗系统中的数字重建影像进行对比,分析等中心在X轴、Y轴、Z轴三个方向上的位移。结果:校准片X轴(左右)、Y轴(头脚)、Z轴(前后)平均误差分别为(0.72±0.57)、(0.41±0.41)、(0.86±0.65)mm;加速器验证片X轴、Y轴、Z轴平均误差分别为(0.77±0.58)、(0.49±0.44)、(0.84±0.62)mm,2组差异无统计学意义(P>0.05)。结论:应用模拟机进行位置校准可以减少定位、摆位中的误差,是一种有效、便捷的位置验证方法。 Objective:To evaluate the value of position verification in intensity modulated radiation therapy(IMRT) by utilizing simulators in head and neck cancer.Methods:Thirty-six cases with head and neck cancer treated with intensity modulated radiotherapy were taken calibration films using simulator for position calibration,and then taken verification films in the same position on the accelerator chip before radiotherapy.Select the typical signs of bone as a reference point for comparison,we analyzed the displacement of iso centers of simulator and accelerator chip calibration and verification films and the digitally reconstructed images from Pinnacle3 treatment system in X axis,Y axis,Z axis.Results:Thirty-six patients with head and neck intensity modulated radiation therapy film X-axis calibration(left),Y-axis(first leg),Z-axis(front) of the average error was(0.72±0.57),(0.41±0.41),(0.86±0.65) mm;accelerator verification films X-axis,Y axis,Z-axis the average error was(0.77±0.58),(0.49±0.44),(0.84±0.62) mm,the difference between two sets did not statistically significant(P0.05).Conclusions:The simulator for position calibration can reduce the positioning error in positioning is an effective and convenient location verification.
出处 《蚌埠医学院学报》 CAS 2012年第1期25-26,共2页 Journal of Bengbu Medical College
关键词 头颈部肿瘤 适形调强放疗 模拟机 摆位误差 head and neck neplasms conformal and intensity modulated radiation therapy simulator set up errors
  • 相关文献

参考文献5

二级参考文献33

  • 1吴云来,赵家成,张雷,李文海,江浩.头颈部肿瘤外照射面罩体位固定技术应用价值的研究[J].实用全科医学,2006,4(5):537-538. 被引量:10
  • 2徐全敬,段诗苗.整体挡铅技术在鼻咽癌放疗中的应用[J].蚌埠医学院学报,2006,31(6):643-644. 被引量:3
  • 3Samuelsson A,Mercke C,Johansson K A. Systematic set-up er rors for IMRT in the head and neck region: effect on dose distribution[J]. Radiother Oncol,2003, 66(3) :303-311. 被引量:1
  • 4Beckham W A, Keall P J, Siebers J V. A fluence convolution method to calculate radiation therapy dose distributions that incorporate random set-up error[J]. Phys Med Biol, 2002,47 (19):3465-3473. 被引量:1
  • 5Stroom J C,de Boer J C,Huizenga H,et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probility[J]. Int J Radiat Oncol Biol Phys, 1999,43(4):905-919. 被引量:1
  • 6Van Herk M,Remeijer P,Rasch C,et al. The probility of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy[J]. Int J Radiat Oncol Biol Phys,2000,47(4) :1121-1135. 被引量:1
  • 7van Lin E N, van der Vight L, Huizenga, et al. Set-up improvement in head and neck radiotherapy using a 3D off line EPID based correction protocol and a customised head and neck sup port[J]. Radiother and Oncol,2003,68(2): 137-148. 被引量:1
  • 8Mock U, Dieckmann K, Wolff U, et al. Portal imaging based definition of the planning target volume during pelvic irradiation for gynecological malignancies[J]. Int J Radiat Oncol Biol Phys,1999,45(1) :227-232. 被引量:1
  • 9Tsai J S, Engler M J, Ling M N, et al. A non-invasive immobiliza tion. system and related quality assurance for dynamic intensity modulated. radiation therapy of intracranial and head and neck disease[J].Int J Radiat Oncol Biol Phys, 1999,43 (2) :455 - 467. 被引量:1
  • 10Wu J, Haycocks T, Alasti H, et al. Positioning errors and prostate motion during conformal prostate radiotherapy using on line isocentre set up verification and implanted prostate markers[J]. Radiother Oncol,2001,61(2) :127-133. 被引量:1

共引文献48

同被引文献38

引证文献3

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部