期刊文献+

浅析毛细管熔点仪测试过程中的影响因素 被引量:1

Analysis of the influencing factors of capillary melting point instrument in the determination process
下载PDF
导出
摘要 利用 MP90毛细管熔点仪对固体样品进行熔点测定实验,从样品处理、操作条件和操作模式三方面进行测试过程中的影响因素分析。结果表明,对于固体试样测试,毛细管的最佳装填高度为3~4 mm;起始温度设置低于物质熔点的标称值前5℃;升温速率越快,熔点测定值越高,数据间的重现性越差;阈值 B 法评定的熔点数值略低于终点 C 法,但测试结果的精密度略高于终点 C 法。 The factors of the sample conditions,operating conditions and operating modes have been analyzed by capillary melting point instrument of MP90.The results show that optimal filling the capillary is about 3 ~4 mm for testing of the solid samples;Starting temperature is set lower than 5 ℃ of the nominal value of the melting point of the material ;The faster heating rate,the higher melting point value,the poorer data reproducibility;The melting valued tested by the threshold value B is slightly lower than Method C endpoint,but the precision of the test results is slightly higher than Method C endpoint.
作者 周丽华 吴颖
出处 《合成技术及应用》 2013年第4期53-56,共4页 Synthetic Technology & Application
关键词 熔点 毛细管 影响因素 melting point capillary influencing factors
  • 相关文献

参考文献4

二级参考文献7

  • 1王昉.药物熔点测定中的DSC实验影响因素研究[J].南京师范大学学报(工程技术版),2004,4(4):58-60. 被引量:15
  • 2北京大学数学力学系数学专业概率统计组.正交设计[M].北京:人民教育出版社,1976.. 被引量:15
  • 3国家药典委员会.中国药典[M].北京:化学工业出版社,2000.87. 被引量:43
  • 4中华人民共和国药典委员会.中国药典.二部.北京:化学工业出版社,2005.附录37. 被引量:1
  • 5[1]刘晓晴,张罗红,李波. 热分析技术在药物分析中的应用进展[R]. 上海: 药物分析杂志编辑部,2003.1-3. 被引量:1
  • 6[3]Wesley Wm Wendlandt. Thermal Analysis[M]. Third Edition. New York: Wiley,1986. 被引量:1
  • 7[6]Breuer K H, Langfelderova H. Thermochimica Acta[M]. New York:Elsevier Science,1982.57,317-329. 被引量:1