移动式电子束术中放疗系统日常质量保证程序的建立及剂量参数稳定性分析 被引量:4

Establishment of daily quality assurance program and analysis of dosimetric characteristics' longterm stability for mobile intraoperative radiation therapy accelerator
下载PDF
导出
摘要 背景与目的:目前移动式电子束术中放疗已成为多种恶性肿瘤重要的辅助治疗手段,本研究旨在建立该系统的日常质量保证(quality assurance,QA)程序并评价其剂量参数的稳定性。方法:QA程序包括安全性及功能性和能量指标及输出剂量的测试,并对2年来质控数据进行分析,评价剂量参数的稳定性。结果:安全性和功能性的测试结果均正常;4、6、9和12 MeV电子束的能量指标分别为(0.666±0.015)mm、(0.839±0.009)mm、(0.781±0.010)mm和(0.724±0.009)mm,输出剂量误差分别为(0.511±0.671)%、(0.278±0.516)%、(0.368±0.532)%和(0.382±0.912)%,均符合相应标准且无随时间变化的趋势。结论:建立的日常QA程序适合于术中放疗系统,且该系统自身的稳定性能满足临床使用要求。 Background and purpose: Currently, intraoperative radiation therapy (IORT) has become the adjuvant therapy of cancer. The study was to establish the daily quality assurance (QA) program and analyze dosiometric characteristics' long-term stability for mobile IORT accelerator. Methods: The QA program of this study included two parts: safety and functionality and energy and output. The two years' QA datasets were acquired and analyzed to investigate the stability of energy and output. Results: All safety and functionality tests passed on a daily basis. The energy index was (0.666±0.015)mm, (0.839±0.009)mm, (0.781 ±0.010)ram, (0.724±0.009)mm and the output dose error was (0.511±0.671)%, (0.278 ± 0.516)%, (0.368 ±0.532)%, (0.382±0.912)% for all energy, respectively. There was no significant time trend in the dosimetric characteristics. Conclusion: The daily QA program is suitable for mobile IORT accelerator. The long-term stability is acceptable for IORT in clinical use.
出处 《中国癌症杂志》 CAS CSCD 北大核心 2014年第1期52-56,共5页 China Oncology
关键词 术中放疗 质量保证 稳定性 Intraoperative radiation therapy Quality assurance Long-term stability
  • 相关文献

参考文献1

二级参考文献7

  • 1Bettar, et al. A New Approach to Intraoperative Radiation Therapy. Aom Journal[J]. 2001,74(4):500-505. 被引量:1
  • 2Felipe A.Calvo, et al. Intraoperative Radiation Therapy First part: Rationale and Techniques. Critical Reviews in Oncology/Hematology[J]. 2006, 59: 106-115. 被引量:1
  • 3M.L.Meurk, et al. The Mobetron: A New Concept for Intraoperative Radiotherapy. The 6th International IORT Symposium [J]. 1996,22- 25. 被引量:1
  • 4Michael, et al. Commissioning of a Mobile Electron Accelerator for Intraoperative Radiotherapy. Journal of Apphed Clinical Medical Physics[J]. 2001,2(3):121-130. 被引量:1
  • 5A.Sam Beddar. Stability of a Mobile Electron Linear Accelerator System for Intraoperative Radiation Therapy. Medical physics[J]. 2005,32 (10): 3128-3132. 被引量:1
  • 6Customer Acceptance Test Procedure, Document: 20444-1: 1-38. 被引量:1
  • 7朱远,张鸿未.术中放疗的现状与未来[J].国外医学(肿瘤学分册),1999,26(3):162-165. 被引量:6

共引文献7

同被引文献29

引证文献4

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部