期刊文献+

CAD/CAE技术在化学发光成像仪设计中的应用研究

Application of CAD/CAE technology in the design of chemiluminescent imaging instrument
下载PDF
导出
摘要 目的:研制一种用于生物芯片、免疫渗滤(IFA)及免疫层析(ICA)等化学发光信号检测的多用途成像仪器。方法:借助计算机辅助设计与辅助工程(CAD/CAE)技术,针对设计的多用途成像仪关键部件进行建模和仿真,对整体调焦定位架构进行分析,得到其承载静强度分布情况;以电荷耦合器件(CCD)摄像头作为热源,对其散热结构进行热-固耦合分析,得到热应力情况,在保证性能的前提下进一步优化结构。结果:通过对整体调焦定位架构和热-固耦合的仿真分析,为改进设计提供理论依据。同时,可减少开发费用及缩短开发周期。结论:CAD/CAE技术是进行产品结构优化设计的有效方法,是设计工作由经验类比方法向科学分析方法转变的重要保证,对于设计工作中提升产品性能、提高设计效率可起到积极的推动作用。 Objective: To develop a multipurpose imaging instrument which can detect chemiluminescent signal of biochips, Immununofiltration assay and Immunochromatographic assay. Methods: With CAD/CAE technology, some key parts undertaldng loads were modeled and analized, including strength analysis and thermal-stress coupled analysis. This ensure to get optimization result. Results: Through simulation analysis, the design level was improved, development costs and development cycle were reduced. Conclusion: Optimization of products with CAD/CAE is an effective design method, which is an important guarantee for the design from the experience analogymethod to the scientific analysis method~ The product's performance arid design efficiency can be iml^roved.
出处 《中国医学装备》 2014年第1期30-33,共4页 China Medical Equipment
关键词 计算机辅助设计 耦合 优化分析 化学发光 Computer aided design Termal-solid coupling Structure optimization Chemiluminescence
  • 相关文献

参考文献15

  • 1王升启.基因芯片技术及应用研究进展[J].生物工程进展,1999,19(4):45-51. 被引量:53
  • 2Kulmala S,Suomi J.Current status of modern analytical luminescence methods[J],Anal Chim Acta, 2003,500(1) .. 21-69. 被引量:1
  • 3Chovin A, Garrigue P, Sojic N.Electrochemilum- inescent detection of hydrogen peroxide with an imaging sensor array[J].Electrochim Acta, 2004,49 : 3751-3757. 被引量:1
  • 4Mayr T,Igel C,Liebsch G,et al. Cross-reactive metal ion sensor array in a micro titer plate format[J]. Anal them, 2003,75(17) : 4389-4396. 被引量:1
  • 5Botchkareva AE,Eremin SA,Montoya A,et al.Development of chemiluminescent ELISAs to DDT and its metabolites in food and environmental samples[J].J Immunol Methods,2003,283(1-2):45-57. 被引量:1
  • 6Marquette CA,Hezard P,Degiuli A,et al.Macro- molecular chemiluminescent complex for enhanced immuno-detection onto microtiter plate and protein biochip[J].Sens Actuators B : Chem, 2006,113: 664-670. 被引量:1
  • 7Kricka LJ. Clinical applications of chemilumin- escence[J]. Anal Chim Acta, 2003,500: 279-286. 被引量:1
  • 8Chouhan RS,Vivek Babu K,Kumar MA,et ahDetection of methyl parathion using immune- chemiluminescence based image analysis using charge coupled device[J].Biosens Bioelectr on,2006,21(7): 1264-1271. 被引量:1
  • 9O'Brien MJ 2nd,Perez-Luna VH,Brueck SR,et al.A surface plasmon resonance array biosensor based on spectroscopic imaging[J]. Biosens Bioelectron,2001,16(1-2):97-108. 被引量:1
  • 10谢友柏.在互联网上实现合作设计[J].航空制造技术,2001,44(2):23-27. 被引量:16

二级参考文献21

共引文献141