期刊文献+

主观因素对SPECT固有泛源均匀性测试影响的探讨 被引量:1

Discussion of subjective factors affecting the intrinsic flood field uniformity of SPECT
下载PDF
导出
摘要 讨论了在点源与探头近距离、探头成角、周围有放射源干扰、采集总计数不足和大剂量点源等主观因素对 SPECT仪的固有泛源均匀性测试结果的影响 ,以指导 The intrinsic flood field uniformity is an important content of SPECT QC, while many factors can contribute to the incorrect measurement of it. For the guarantee of SPECT QC, this paper will discuss the subjective factors affecting the measurements of the SPECT intrinsic flood field uniformity, such as a close distance between radiosource and detector, radiation shooting detector at an angle, the interference of radiation coming from around, insufficient total counts, higher dose of the radiosource etc.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2000年第4期312-314,共3页 Nuclear Electronics & Detection Technology
关键词 SPECT 主观因素 固有泛源 均匀性测试 SPECT uniformity subjective factor affection
  • 相关文献

参考文献4

二级参考文献4

  • 1袁爱娜 赵惠扬.-[J].中华核医学杂志,1994,(10):58-60. 被引量:1
  • 2伊大一 刘自来 等.-[J].中华核医学杂志,1995,15:199-199. 被引量:1
  • 3伊大一,中华核医学杂志,1995年,15卷,199页 被引量:1
  • 4袁爱娜,中华核医学杂志,1994年,14卷,10期,58页 被引量:1

共引文献6

同被引文献14

引证文献1

二级引证文献2

相关作者