双探头SPECT仪质量控制方法探讨 被引量:18

Quality Control of Dual-head SPECT
下载PDF
导出
摘要 目的:探讨双探头SPECT仪质量控制中主要性能指标的可检测方法及意义。方法:用符合NEMA标准的自制99mTc点源、四象限铅栅模型和Jaszczak断层模型分别对仪器能峰、能量分辨率、CFOV固有积分均匀性、固有空间分辨率和固有线性、系统灵敏度和断层均匀性、分辨率等指标进行测试。结果:2007-06—2009-06所测得的质控结果均符合要求,仪器处于正常状态。结论:合理的质控方法可以有效保证仪器的正常运行,提供优质的图像。 Objective To investigate the testing methods for chief performances in quality control(QC)of dual-head SPECT.Methods With ^99mTc point source,bar phantom and Jaszczak phantom in accordance with the NEMA standard,performances testing was done as to energy peak position,energy resolution,CFOV intrinsic integral uniformity,intrinsic spatial resolution and linearity,system sensitivity,and ECT uniformity,resolution etc.Results All the QC results from June,2007 to June,2009 were acceptable.Conclusion Reasonable QC methods can keep the SPECT machine in good condition,and provide good image.
出处 《医疗卫生装备》 CAS 2010年第7期116-116,132,共2页 Chinese Medical Equipment Journal
关键词 SPECT 质量控制 分辨率 均匀性 SPECT quality control resolution uniformity
  • 相关文献

参考文献7

二级参考文献37

  • 1张峰,程木华,郑子梅,张勇,吴春兴,陈谊.探讨符合线路SPECT/CT的日常质量控制与质量保证[J].现代医学仪器与应用,2007,0(3):42-45. 被引量:27
  • 2Stark SA, Ohlsson J, Carlsson S. An evaluation of reconstruction techniques and scatter correction in bone SPECT of the spine. Nuclear Medicine Communications, 2003, 24,565- 570. 被引量:1
  • 3Blocklet D, Seret A, Popa N, et al. Maximum - likelihood reconstruction with ordered subsets in bone SPECT. The Journal of Nuclear Medicine, 1999, 40: 1978- 1984. 被引量:1
  • 4Case JA, Lincho R, Michael A, et al. Bone SPECT of the spine: a comparison of attenuation correction techniques. The Journal of Nuclear Medicine, 1999, 40: 604- 613. 被引量:1
  • 5Wells RG, Farncombe T, Chang E, et al. Reducing bladder arti- facts in clinical pelvic SPECT Images. The Journal of Nuclear Medicine, 2004, 45: 1309-1314. 被引量:1
  • 6贺小红,贺小红.NEMA标准的参考资料、概念定义和测试设备[J].国外医学(放射医学核医学分册),1997,21(2):49-51. 被引量:5
  • 7李小华.用于SPECT的单晶体闪烁相机测试方法:性能标准[J].国外医学(放射医学核医学分册),1997,21(2):61-62. 被引量:1
  • 8International Electrotechnical Commission. IEC 789. Characteristics and test conditions of radionuclide imaging devices ; anger type gamma cameras. Switzerland : IEC, 1992. 被引量:1
  • 9International Electrotechnical Commission. IEC 61675-2. Radionuclide imaging devices-characteristics and test conditions-part 2 : Single photon emission computed tomographs. Switzerland : IEC, 1998. 被引量:1
  • 10International Atomic Energy Agent. IAEA TECDOC-602. Quality control of nuclear medicine instruments 1991. Vienna: IAEA, 1991. 被引量:1

共引文献55

同被引文献109

引证文献18

二级引证文献63

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部