期刊文献+

在线离子交换预富集–FLAAS测定环境水样中的铬(Ⅵ) 被引量:1

Determination of Cr( Ⅵ ) in Water Sample After Preconcentration by On-line Ion Exchange FLAAS
下载PDF
导出
摘要 采用在线离子交换预富集–火焰原子吸收光谱法(FLAAS)测定环境水样中痕量铬(Ⅵ)。通过试验考察样品溶液pH、洗脱剂浓度、离子交换树脂用量及共存离子对离子交换树脂富集效果的影响。结果表明,当交换树脂用量为0.50 g,样品溶液pH值为6.0时,用0.60 mol/L盐酸–10%抗坏血酸进行洗脱具有良好效果。铬(Ⅵ)的质量浓度在0~20.0μg/L之间与吸光度呈良好的线性关系,线性相关系数大于0.9998。该方法用于在线分离和富集环境水样中的铬(Ⅵ),灵敏度提高了100倍,加标回收率为96%~104%。 Trace amount of Cr ( Ⅵ ) was determined by using on-line ion-exchange column preconcentration flame atomic absorption spectroscopy system (IE-FLAAS). The effects of experimental variables such as the pH of sample, the concentration of eluent, the content of ion-exchange resin and coexisting ions on the recovery of Cr ( VI ) were inspected. The optimum operation conditions were as follows:the content of ion-exchange resin was 0.50 g, the pH of solution was 6.0, the eluent was 0.60 mol/L HCI-10% ascorbic acid. The mass concentration of Cr ( VI ) was linear with absorbance in the range of 0-20.0 μ g/L, the correlation coefficient was more than 0.999. The proposed method was used for on-line separation and preconcentration of Cr ( VI ) in water sample, the sensitivity of IE-FLAAS was 100 times higher than that of FLAAS and recovery was in the range of 96%-104%.
出处 《化学分析计量》 CAS 2013年第1期27-29,共3页 Chemical Analysis And Meterage
基金 重大科学仪器设备开发专项(2011YQ030124 2011YQ060100)
关键词 离子交换分离 预富集 火焰原子吸收光谱法 铬(Ⅵ) ion-exchange separation preconcentration flame atomic absorption spectroscopy (FLAAS) Cr ( VI )