摘要 目的研究Diazyme和北京九强2家公司的糖化血红蛋白检测试剂盒的可比性。方法收集佛山市顺德区龙江人民医院检验科当日的高值(大于6%)和正常值(3%~6%)的血清标本,在生化分析仪上,分别用2家公司的试剂盒检测糖化血红蛋白,统计分析其相关系数。结果 2家公司的糖化血红蛋白检测值高值的相关系数r2为0.98,正常值的相关系数为0.97。结论 2家公司的糖化血红蛋白检测值有良好的相关性。 Objective To research the comparison of HbAlc assay Kit of two companies. Methods The serums of the high(〉6%)and normal value(3%- 6%) were collected in the laboratory section. Biochemical analyzer was used to test the HbA1c with two kits from Diazyme and Jiuqiang, respectively. Results were recorded and analyzed. Results The correlation coefficient of the high and normal values were 0.98 and 0.97. Conclusion A good correlation of HbAlc has been shown between Diazyme and Jiuqiang.
出处 《检验医学与临床》 CAS 2012年第23期2935-2936,共2页 Laboratory Medicine and Clinic
关键词 糖化血红蛋白 检测试剂盒 血清 相关系数 glycosylated hemoglobin assay kit serum correlation coefficient
  • 相关文献

参考文献8

二级参考文献25

  • 1沈钰如.长期糖基化终末产物与糖尿病并发症[J].国外医学(内分泌学分册),1995,15(2):86-87. 被引量:8
  • 22,叶应妩,王毓三主编.全国临床检验操作规程.南京:东南大学出版社,1990: 被引量:1
  • 34,美国ABBOTT公司IMX.Glycated Hemoglobin.Note Changes Highlighted.October,1997: 被引量:1
  • 45,Wilson DH et al.Fully Automated Assay of Glycohemoglobin with the IMX Analyzer:Novel Approaches for Separation and Detection.Clin Chem,1993;39:20907 被引量:1
  • 5中华医学会糖尿病学会.中国2型糖尿病防治指南(2007年).中华内分泌代谢杂志,2008,24(2):2-4. 被引量:2
  • 6Flii ckiger R, Mortensen HB. Glycated haemoglobins [J]. J Chromatogr, 1988,429:279-292. 被引量:1
  • 7Thevarajah TM, Nani N, Chew YY. Performance evaluation of the Arkray Adams HA-8160 HbAlc analyser [J]. Malays J Pathol, 2008,30(2):81 -86. 被引量:1
  • 8John WG, Mosca A, Weykamp C, et al. HbAlc stan- dardisation : history, science and politics [ J ]. Clin Biochem Rev, 2007, 28(4):163 -168. 被引量:1
  • 9Hoelzel W, Weykamp C, Jeppsson JO, et al. IFCC reference system for measurement of hemoglobin Alc in hu- man blood and the national standardization schemes inthe United States, Japan, and Sweden: a method-com- parison study [ J ]. Clin Chem, 2004, 50 ( 1 ) : 166 - 17d. 被引量:1