摘要
目的检测某品牌移动式空气净化消毒机对室内空气中细菌的清除作用,以验证其对空气消毒的效果。方法《消毒技术规范》(2002)2.1.3.5空气消毒效果鉴定现场试验。结果该空气净化消毒机对医院ICU病房中的空气消毒作用30min,室内空气中自然菌的消亡率分别为91.45%(低档)、99.38%(高档)。结论该移动空气净化消毒机对医院Ⅱ类环境空气中自然菌的清除符合《消毒技术规范》(2002)的要求。
Objective Test the disinnfection of bacteria in indoor air for air disinfection machine,to verify its effectiveness to air disinfection.Methods According to "Technical Standard For Disinfection"(2002) 2.1.3.5 "Identification of air disinfection on the spot".Results After the air disinfection machines disinfect air 30 mins in the ICU of the hospital,the disappearance rates of natural bacteria in indoor air were 91.45%(low rank) and 99.38%(high rank) respectively.Conclusion The clearance of the natural bacteria in air of hospital class Ⅱ environment for the air disinfection machine meets the requirement of " Technical Standard For Disinfection "(2002).
出处
《中国医药指南》
2012年第26期28-29,共2页
Guide of China Medicine
关键词
消毒
空气
Disinfection
Air