摘要
介绍了STANNO Metal 9000型镀锡量分析仪的工作原理,并对在使用过程出现的仪器自测无法通过、电位检测异常以及电解槽等故障进行了分析,并提出了处理措施。
Failures in the use of Stanno Metal 9000 tin not passed, abnormality in voltage measurement, abnormal coating analyzer are analyzed, such as self-test electrolytic cell, etc. Methods are also presented for their elimination. The result presents an important reference for solving similar problems.
出处
《武汉工程职业技术学院学报》
2012年第3期28-31,共4页
Journal of Wuhan Engineering Institute