BC-5500与SYSMEX XS-800i血液分析仪性能评价 被引量:1

The Evaluation of Performance of BC-5500 Plus and SYSMEX XS-800i Blood Cells Analyzer
下载PDF
导出
摘要 目的:评价BC-5500与SYSMEX XS-800i全自动血细胞分析仪性能,以保证不同仪器测定结果的准确性和一致性。方法:按照国际血液学标准化委员会(ICSH)有关规定,对BC-5500和SYSMEX 800i全自动血细胞分析仪从精密度、携带污染率、线性、及ABX-120全自动血细胞分析仪相关性及白细胞分类准确性等方面进行评价。结果:BC-5500及SYSMEX 800i的批间、批内精密度及总精密度变异系数(CV%)均在可接受范围内;各分析参数的携带污染率均<1.0%;与ABX-120对比分析,各项参数相关性良好;白细胞分类准确性符合要求。结论:BC-5500及SYSMEX 800i血液分析仪性能良好,检测结果准确可靠,在临床应用中可以互换。 Objective: To evaluate the main performance of BC - 5500 plus and SYSMEX XS - 800i automatic blood cells analyzer, to ensure the two instrument the veracity of test results and consistency. Methods :According to international hematology standardization commission (ICSH) with the related regulations, the performances of BC -5500 and SYSMEX XS -800i hematology analyzer were evaluated in regards regard to precision, carry - over rate, lineality, thecorrelation between the ABX -120 Heamatology analyzer in detectin results and the accuracy of WBC classification. Results: The CV value of with - and between - Batch precision and that of the total precision of BC - 5500 and SYSMEX 800i were all within the designed range. The carry - over rate was lower than 1.0%. The correlation between ABX - 120 was satisfactory. The accuracy of WBC classification meets the require- ment. Conclusions: The performances of MINRAY BC - 5500 and SYSMEX XS - 800i hematology an- alyzer have proved to be accurate and reliable in test results.
出处 《航空航天医学杂志》 2012年第2期132-134,共3页 Journal of Aerospace medicine
关键词 血液分析仪 精密度 携带污染率 相关性 准确性 Blood cells analyzer Precision Carry - over rate Correlation Accuracy
  • 相关文献

参考文献5

  • 1International Council for Standardization in He-matology,Guidelines for the evaluation of bloodcell analyzer including those used for differentialleukocyte and reticulocyte counting and cellmarker application.Prepared by the ICSH expertpanel on cytometry[J].Clin Lab Haemat,1994,16(2):157-174. 被引量:1
  • 2CLIA88 USA.Clinacal Laboratory ImprovementAmendment[S].1988. 被引量:1
  • 3藤本敬二.血细胞分析仪的溯源及国际参考方法[J].中华检验医学杂志,2004,27(1):53-54. 被引量:56
  • 4叶应妩,王毓三,申子瑜主编..全国临床检验操作规程 第3版[M].南京:东南大学出版社,2006:1034.
  • 5丛玉隆.血液分析仪进展及临床应用[M].1989:49. 被引量:2

二级参考文献6

  • 1Lewis SM.Harmonization and standardization of the blood countthe role of the ICSH cytometry panel-[].Sysmex J Int.1999 被引量:1
  • 2Bain,Barbara J.Detecting erroneous blood counts, chapter 4 (132146) of blood cells -a practical guide- 2nd Edition, Blackwell Science[].United Kingdom.1995 被引量:1
  • 3Cornbleet J.Spurious results from automated hematology cell counters[].Laboratory Medicine.1983 被引量:1
  • 4ICSH HGB Expert Panel.Recommendation for reference method for haemoglobinometry in human blood and specifications for international haemiglobincyanide standard[].Journal of Clinical Pathology.1996 被引量:1
  • 5ICSH Cytometry Panel.Platelet counting by the RBC/patelet Ratio Method-A reference Method-[].American Journal of Clinical Pathology.2001 被引量:1
  • 6ICSH Cytometry Panel.Recommendation for Reference method for the packed cell volume[].Laboratory Hematology.2001 被引量:1

共引文献56

同被引文献12

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部