期刊文献+

NPD与FPD气相色谱法测定水中的黄磷 被引量:6

Determination of Yellow Phosphorus in Water by NPD and FPD Gas Chromatography
下载PDF
导出
摘要 实验采用优化条件,对甲苯萃取水和废水中黄磷进行测定。萃取液经气相色谱仪毛细色谱柱分离后,氮磷检测器(NPD)或火焰光度检测器(FPD,带磷滤光片)检测,根据色谱峰的保留时间定性,外标法定量。使用NPD检测器分析时,本方法检出限为0.04μg/L;使用FPD检测器分析时,本方法检出限为0.1μg/L。 Experiments using optimized conditions determined yellow phosphorus in toluene extraction water and wastewater.After extracted liquid by gas chromatography capillary column separation,detected by nitrogen and phosphorus detector(NPD) or flame photometric detector(FPD with phosphorus filter),according to the chromatographic peak retention time qualitative,quantified by external standard.Analysis of the NPD detector,the method detection limit was 0.04 μg/L;FPD detector,the detection limit was 0.1 μg/L.
出处 《干旱环境监测》 2012年第1期1-4,共4页 Arid Environmental Monitoring
关键词 NPD FPD 气相色谱 黄磷 NPD FPD gas chromatography yellow phosphorus
  • 相关文献

参考文献5

二级参考文献2

  • 1李浩春.气相色谱法[M].北京:科学出版社,1998.248. 被引量:23
  • 2《水和废水监测分析方法指南》编委会.水和废水监测分析方法指南(下册)[M].北京:中国环境科学出版社,1997.25-25. 被引量:2

共引文献417