期刊文献+

数字X光影像仪光机结构设计

Structure Design of Computed Radiography
下载PDF
导出
摘要 设计了数字X光影像仪的光机结构。针对数字X光影像仪的特点和指标要求,确定了光机结构总体布局。介绍了数字X光影像仪的关键部分片盒、输片机构、激光扫描系统、荧光读取系统和消光系统的具体结构,并对输片传动机构进行了输片精度分析。检测结果表明,扫瞄轴晃动1.2″,输片精度0.015mm。 The structure design of computed radiography was introduced.The structure collectivity layout was ascertained in allusion to characteristics and target request of computed radiography.The key structure of IP box,transporting device,laser scanning systems,fluorescence collecting systems and extinction systems were introduced.IP-transport precision was analyzed.The measurement results showed that the scanning axis sway was 1.2″ and IP-transport precision was 0.013 mm.
作者 孟庆华
出处 《光机电信息》 2011年第8期32-36,共5页 OME Information
关键词 数字X光影像仪 结构设计 输片精度 扫瞄轴晃动 computed radiography structure design IP-transport precision scanning axis sway
  • 相关文献

参考文献9

二级参考文献20

  • 1刘欣悦,黄廉卿.利用多分辨率直方图特征分类数字X光乳腺图像[J].光学精密工程,2006,14(2):327-332. 被引量:10
  • 2杨皓明,王灵杰,冯秀恒.一种新型复合摄像物镜的设计[J].光学精密工程,2007,15(4):486-491. 被引量:5
  • 3[美]Peter G Viscarola等 新智工作室(译).Windows NT与Windows 2000设备驱动及开发[M].北京:电子工业出版社,2000.. 被引量:1
  • 4熊光楠.二阶杂质正离子对F心的干扰[J].发光学报,1999,20:135-138. 被引量:1
  • 5[2]SODERMAN M, PICARD C, ERICSON K. An algorithm for correction of distortion in stereotaxic digital subtraction angiography[J]. Neuroradiology, 1998,40: 277-282. 被引量:1
  • 6Brabdt Michael.the design and peformance characteristics of collecter high efficiency detection of photosimulable phospher emissions Prec SPIE, 1993,1896:288 - 295. 被引量:1
  • 7张以谟.应用光学[M].北京:机械工业出版社,1984. 被引量:6
  • 8CARLSSON T E, ORNHED H. Evaluation of digital flash X-ray images produced using computed radiography[J]. SPIE, 2005, 5580:842-850. 被引量:1
  • 9MURTRY E K. Laser scanning lens configurations[J]. SPIE, 1990,1333:145-158. 被引量:1
  • 10THIBAULT S. F-Sin(theta) lens system and applications[J]. SPIE, 2001, 4441:98-105. 被引量:1

共引文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部