乳腺DR成像普通模式和精细模式成像性能对比 被引量:1

Experiments on Choice of Exposure Mode for Digital Mammography
下载PDF
导出
摘要 目的对数字乳腺X射线摄影的两种不同曝光模式(精细模式和普通模式)成像性能进行评价。方法采用乳腺成像体模,选择不同的阳极滤过器组合,分别在两种自动曝光模式下成像,测量图像信噪比和腺体吸收剂量,并引入图像品质因数综合评价成像性能的好坏。结果精细模式图像信噪比要高于普通模式,但使腺体吸收剂量增加,因而品质因数低于普通模式。结论在临床使用中必须充分考虑临床实际需要来选择相应的成像技术,在保证临床图像质量的同时尽量降低受检者的辐射风险。 Objective To compare the imaging characteristic of two automatic exposure modes(H mode and D mode).Methods contrast and resolution mammography phantom was used to simulate human breast tissue and three different combinations of anode/filter material were employed under the two automatic exposure modes.The image signal to noise ratio(SNR) and mean glandular dose(MGD)were measured.The figure of merit(FOM) was used to evaluate the imaging performance of digital mammography.Results the SNR and MGD were higher in H mode than D mode,but the FOM was lower in H mode than D mode.Conclusion the Optimization for digital mammography is very important for clinical use in order to keep the good image quality and reduce the radiation dose.
出处 《中国医疗设备》 2011年第5期51-53,共3页 China Medical Devices
基金 国家自然科学基金项目(30870751) 北京市卫生系统高层次卫生技术人才培养计划项目(2009-3-57)
关键词 数字乳腺X线摄影 平板探测器 信噪比 平均腺体吸收剂量 品质因数 digital mammography flat panel detector signal to noise ratio mean glandular dose figure of merit
  • 相关文献

参考文献10

  • 1Van Ongeval C,Bosmans H,Van Steen A,et al.Evaluation of the diagnostic value of a computed radiography system by comparison of digital hard copy images with screen-film mammography:results of a prospective clinical trial[J].Eur Radiol, 2006,16:1360-1366. 被引量:1
  • 2Young KC,Cook JJH,Oduko JM.Use of the European Protocol to Optimise a Digital Mammography System[C]. In:Proceedings of the 8th International Workshop on Digital Mammography,2006:362-369. 被引量:1
  • 3Uchiyama N, Moriyama N, Kitagawa M. Clinical Optimization of Filters in Direct a-Se FFDM (Full Field Digital Mammog- raphy) System[J]. IWDM 2006, LNCS4046,315-323,2006. 被引量:1
  • 4Saunders RS,Samei E,Jesneck JL,et al.Physical characterization of a prototype selenium-based full field digital mammography detector[J].Med Phys,2005,32:588-599. 被引量:1
  • 5Young KC.Breast dose surveys in the NHSBSP:software and instruction manual, Version 2.0[M]. NHSBSP Report 04/05, Guildford,2004. 被引量:1
  • 6Dance DR,Skinner CL,Young KC,et al.Additional factors for estimation of mean glandular breast dose using the UK mammography dosimetry protocol[J].Phys Med Biol 2000,45: 3225-3240. 被引量:1
  • 7Perry N, Broeders M, de Wolf C, et al. European guidelines for quality assurance in breast cancer screening and diagnosis,4th edition[M].European Communities 2006,Luxembourg: chapter2. 被引量:1
  • 8Toroi P,Zanca F,Young KC,et al.Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system[J].Eur Radiol 2007,17:2368-2375. 被引量:1
  • 9白玫,彭明辰.X线CT成像质量评估中的调制传递函数的测试与分析[J].医疗设备信息,2006,21(5):11-13. 被引量:15
  • 10徐莉,梁泉,杨刚.婴幼儿DR胸片的质量控制和优势[J].中国医疗设备,2009,24(12):128-129. 被引量:9

二级参考文献12

  • 1全国放射科QA,QC学术研讨会纪要[J].中华放射学杂志,1993,27(2):134-138. 被引量:134
  • 2Garmer M,Henning SPJager H,et al Digitai radiography versus conventional in chest imaging diagnostic performance of a large area silicon flat-detector in a clinic CT-controlled study[J]. AJR,2000, 174(1):75. 被引量:1
  • 3Niklason LT,Chan HP,Casade PN,et al.Ppontablechest imahing:comparison of storage phosphor digital,asymmetric sceen-film,andconventional sceen-film system[J].radiology, 1993,186:387-393. 被引量:1
  • 4Goodenough D,Atkins F,Dyer S.Automated techniques for quality assurance of radiological image modalities.SPIE.Image Capture,Formatting,and Display,1991,1444:87-99. 被引量:1
  • 5Dainty JC,Shaw R.Image Science[M].New York:Academic Press,1974. 被引量:1
  • 6Goodenough D,Atkins F.Theoretical and practical aspects of automated quality assurance approaches,particularly for CT[C].Proceedings of the ICR 89,editors Silve,et.Al,1990. 被引量:1
  • 7Fujita H,Tsai D-Y,Itoh T,et al.A simple method for determining the modulation transfer function in digital radiography[J].IEEE Transactions on Medical Imaging,1992,11:34-39. 被引量:1
  • 8Cunningham IA,Reid BK.Signal and noise in modulation transfer function determinations using the slit,wire,and edge techniques[J].Medical Physics,1992,19:1037-1044. 被引量:1
  • 9Samei E,Flynn MJ,Reimann DA.A method for measurement the presampled MTF of digital radiographic systems using an edge test device[J].Medical Physics,1998,25:102-113. 被引量:1
  • 10王西端,陈方,侯钦森,聂存伟,张兴华,王洪春.婴幼儿CR胸部摄影综合技术分析和影像质量评价[J].中国医疗设备,2009,24(4):94-95. 被引量:5

共引文献22

同被引文献11

  • 1Van Ongeval C, Bosmans H, Van Steen A, et al. Evaluation of the diagnostic value of a computed radiography system by comparison of digital hard copy images with screen-film mammography: results of a prospective clinical trial[J]. Eur Radiol,2006,16:1360-1366. 被引量:1
  • 2Saunders RS, Samei E, Jesneck JL, et al. Physical characterization of a prototype selenium-based full field digital mammography detector [J].Med Phys,2005,32:588-599. 被引量:1
  • 3Doyle P, Martin CJ, Gentle D. Application of contrast-to-noise ration in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations[J]. Phys Med Biol,2006, 51:2953-2970. 被引量:1
  • 4Gingold EL, Xizeng W, Barnes GT. Contrast and dose with Mo-Mo, Mo-Rh and Rh-Rh target-filter combinations in mammography[J]. Radiology,1995,195:639-644. 被引量:1
  • 5Dance DR, Thilander Klang A, Sandborg M, et al. Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study[J]. BrJ Radiol,2000,73:1056-1067. 被引量:1
  • 6Fahfig R,Rowlands JA,Yaffe MJ.X-ray imaging with amorphous selenium:Optimal spectra for digital mammography[J].Med Phys,1996,23:557-567. 被引量:1
  • 7Young KC. Breast dose surveys in the NHSBSP: software and instruction manual,Version 2.0[M].NHSBSP Report 04/05, Guild ford,2004. 被引量:1
  • 8Dance DR, Skinner CL, Young KC, et al. Additional factors for estimation of mean glandular breast dose using the UK mammography dosimetry protocol[J]. Phys Med Biol,2000, 45:3225-3240. 被引量:1
  • 9Perry N,Broeders M,de Wolf C,et al.European guidelines for quality assurance in breast cancer screening and diagnosis,4th edition[M]. European Communities,2006,Luxembourg: chapter 2. 被引量:1
  • 10Toroi P,Zanca F,Young KC,et al.Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system[J]. Eur Radiol,2007,17:236B-2375. 被引量:1

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部